English 37
Humidifier performance
The following settings have been tested at 71.6°F (22°C) ambient temperature:
Mask Pressure
cm H
2
O
RH output %
Nominal system output AH
1
, BTPS
2
Setting 4
Setting 8
Setting 4
Setting 8
4 85 100
6 >10
10 85 100 6 >10
20 85 90 6 >10
1
AH - Absolute Humidity in mg/L
2
BTPS - Body Temperature Pressure Saturated
Air tubing
Air tubing
Material
Length
Inner diameter
ClimateLineAir
Flexible plastic and electrical components
6'6" (2 m)
0.6" (15 mm)
SlimLine
Flexible plastic
6' (1.8 m)
0.6" (15 mm)
Standard
Flexible plastic
6'6" (2 m)
0.75" (19 mm)
Heated air tubing temperature cut-out:
106°F (
41°C)
Notes:
The manufacturer reserves the right to change these specifications without notice.
The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should
not be fitted to the mask.
Do not use electrically conductive or antistatic air tubing.
The temperature and relative humidity settings displayed are not measured values.
Displayed values
Value Range Display
resolution
Pressure sensor at air outlet:
Mask pressure
4–20 cm H
2
O
0.1 cm H
2
O
Flow derived values:
Leak
0–120 L/min
1 L/min
Value Accuracy
Pressure measurement
1
:
Mask pressure
2
±[0.5 cm H
2
O + 4% of measured value]
Flow and flow derived values
1
:
Flow
±6 L/min or 10% of reading, whichever is greater, at 0 to 150 L/min positive flow
Leak
2
±12 L/min or 20% of reading, whichever is greater, 0 to 60 L/min
1
Results are expressed at ATPD (Ambient Temperature and Pressure, Dry).
2
Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min.
Pressure accuracy
Maximum static pressure variation at 10 cm H
2
O according to ISO 17510-1:2007
Standard air tubing
SlimLine air tubing
Without humidification
± 0.5 cm H
2
O
± 0.5 cm H
2
O
With humidification
± 0.5 cm H
2
O
± 0.5 cm H
2
O
Summary of Contents for AirSense S10