English 23
Reprocessing the air tubing
Disconnecting
1.
Hold the cuff of the air tubing and gently pull it away from the device.
2.
Hold both the cuff of the air tubing and the swivel of the mask, then gently pull apart.
Decontaminating
Before the disinfection process, each component must be cleaned and rinsed so no visible
contamination is present.
1.
Clean all components with a soft bristled brush for one minute while soaking in detergent
solution (see table below). Pay particular attention to all crevices and cavities.
2.
Run the detergent solution through the air tubing repeatedly until no contamination is visible.
3.
Thoroughly rinse each component according to the detergent manufacturer's instructions.
ResMed has tested the following detergents according to the manufacturer’s instructions:
Detergent Water
temperature
SlimLine ClimateLineAir Standard
Alconox
™
(diluted at 1%)
Hot water (approx 140°F or 60°C)
Warm water (approx 113 to 140°F or 45 to 60°C)
Room temperature water (approx 70°F or 21°C)
Neodisher MediZym
™
(diluted
at 2.0%)
Warm water (approx 113 to 140°F or 45 to 60°C)
High level thermal disinfection
Part
Validated number of cycles
Hot water: 167°F (75°C) for 30 minutes.
SlimLine 20
ClimateLineAir 10
Standard 100
1.
Immerse the air tubing in a water bath.
Take care that no air bubbles are trapped inside the air tubing.
2.
Increase the water bath temperature to 167ºF (75°C) for 30 minutes. Higher temperatures may
damage the tubing.
3.
Air dry out of direct sunlight and/or heat.