English 35
Leave the water tub to cool for ten minutes before handling to allow the water to cool and
to make sure that the water tub is not too hot to touch.
Make sure that the water tub is empty before transporting the device.
Technical specifications
90W power supply unit
AC input range:
100–240V, 50–60Hz 1.0–1.5A, Class II
115V, 400Hz 1.5A, Class II (nominal for aircraft use)
DC output:
24V
3.75A
Typical power consumption:
53W (57VA)
Peak power consumption:
104W (108VA)
Environmental conditions
Operating temperature:
+41°F to +95°F (+5°C to +35°C)
Note:
The air flow for breathing produced by this therapy
device can be higher than the temperature of the room.
Under extreme ambient temperature conditions
(104°F/40°C) the device remains safe.
Operating humidity:
10 to 95% relative humidity, non-condensing
Operating altitude:
Sea level to 8,500' (2,591 m); air pressure range 1013 hPa to
738 hPa
Storage and transport temperature:
-4°F to +140°F (-20°C to +60°C)
Storage and transport humidity:
5 to 95% relative humidity, non-condensing
Electromagnetic compatibility
The AirSense 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-
2:2007, for residential, commercial and light industry environments. It is recommended that mobile communication devices
are kept at least 1 m away from the device.
Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on
www.resmed.com, on the Products page under Service and Support.
IEC 60601-1:2005 classification
Class II (double insulation), Type BF, Ingress protection IP22.
Sensors
Pressure sensor:
Internally located at device outlet, analog gauge pressure
type, -5 to +45 cm H
2
O
Flow sensor:
Internally located at device inlet, digital mass flow type, -70
to +180 L/min
Maximum single fault steady pressure
Device will shut down in the presence of a single fault if the steady state pressure exceeds:
30 cm H
2
O for more than 6 sec or 40 cm H
2
O for more than 1 sec.
Sound
Pressure level measured according to ISO 17510-1:2007 (CPAP mode):
SlimLine:
26.6 dBA with uncertainty of 2 dBA
Standard:
26.6 dBA with uncertainty of 2 dBA
Power level measured according to ISO 17510-1:2007 (CPAP mode):
SlimLine:
34.6 dBA with uncertainty of 2 dBA
Standard:
34.6 dBA with uncertainty of 2 dBA
Declared dual-number noise emission values in accordance with ISO 4871:1996.