Sound
Pressure level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine:
25 dBA with uncertainty of 2 dBA
Standard:
25 dBA with uncertainty of 2 dBA
SlimLine or Standard and humidification:
27 dBA with uncertainty of 2 dBA
Power level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine:
33 dBA with uncertainty of 2 dBA
Standard:
33 dBA with uncertainty of 2 dBA
SlimLine or Standard and humidification:
35 dBA with uncertainty of 2 dBA
Declared dual-number noise emission values in accordance with ISO 4871:1996.
Physical - device and water tub
Dimensions (H x W x D):
4.57" x 10.04" x 5.91"
(116 mm x 255 mm x 150 mm)
Air outlet (complies with ISO 5356-1:2015):
22 mm
Weight (device and standard water tub):
44 oz (1248 g)
Weight (device and cleanable water tub):
44 oz (1248 g)
Housing construction:
Flame retardant engineering thermoplastic
Water capacity:
To maximum fill line 380 mL
Standard water tub - material:
Injection molded plastic, stainless steel and silicone seal
Cleanable water tub - material:
Injection molded plastic, stainless steel and silicone seal
Temperature
Maximum heater plate:
154°F (68°C)
Cut-out:
165°F (74°C)
Maximum gas temperature:
≤
106°F (
≤
41°C)
Air filter
Standard:
Material: Polyester non woven fiber
Average arrestance: >75% for ~7 micron dust
Hypoallergenic:
Material: Acrylic and polypropylene fibers in a polypropylene
carrier
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron
dust
Aircraft use
ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M) for all phases of air travel.
Wireless module
Technology used:
CDMA (USA only)
4G (USA and Canada only)
3G
2G (all regions except USA and Canada)
Frequencies
CDMA (850/1900 MHz)
4G (700/850/1900 MHz)
3G (850/900/1700/1800/1900/2100 MHz)
2G (850/900/1800/1900 MHz)
Max RF power output
CDMA 24.5 dBm
4G 23.0 dBm
3G 24.0 dBm
2G 33.0 dBm