FCC ID: 2ACHL-AIR104G, 2ACHL-AIR103G
IC: 9103A-AIR104G, 9103A-AIR103G
The AirSense 10 device complies with FCC Rules and Industry Canada rules.
The AirSense 10 device should be used at a minimum distance of 0.8" (2 cm) from the body during operation.
Additional information regarding the FCC Rules and IC compliance for this device can be found on
www.resmed.com/downloads/devices
Operating pressure range
AutoSet, AutoSet For Her, Elite, CPAP
4 to 20 cm H
2
O
Supplemental oxygen
Maximum flow:
4 L/min
Pneumatic flow path
1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. Water tub
7. Device
8. Inlet filter
Design life
Device, power supply unit:
5 years
Cleanable water tub:
2.5 years
Standard water tub, air tubing:
6 months
General
The patient is an intended operator.
Humidifier performance
Mask Pressure
cm H
2
O (hPa)
RH output % at
63°F (17°C) ambient
temperature
RH output % at
72°F (22°C) ambient
temperature
Nominal system output AH
1
, BTPS
2
Setting 4
Setting 8
Setting 4
Setting 8
4
85
100
6
>10
10
85
100
6
>10
20
85
90
6
>10
1
AH - Absolute Humidity in mg/L
2
BTPS - Body Temperature Pressure Saturated