20
Pressure accuracy
Maximum static pressure variation at 10 cm H
2
O (10 hPa) according to
ISO 80601-2-70:2015
Standard air tubing
SlimLine air tubing
Without humidification
± 0.5 cm H
2
O (± 0.5 hPa)
± 0.5 cm H
2
O (± 0.5 hPa)
With humidification
± 0.5 cm H
2
O (± 0.5 hPa)
± 0.5 cm H
2
O (± 0.5 hPa)
Maximum dynamic pressure variation according to ISO 80601-2-70:2015
Device without humidification and Standard air tubing / Device with humidification and Standard air tubing
Pressure [cm H
2
O (hPa)]
10 BPM
15 BPM
20 BPM
4
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
8
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
12
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
16
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
20
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
Device without humidification and SlimLine air tubing / Device with humidification and SlimLine air tubing
Pressure [cm H
2
O (hPa)]
10 BPM
15 BPM
20 BPM
4
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
8
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
12
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
16
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
20
0.5 / 0.5
0.5 / 0.5
0.8 / 0.8
Flow (maximum) at set pressures
The following are measured accordingly to ISO 80601-2-70:2015 at the end of the specified air tubing:
Pressure
AirSense 10 and
Standard
AirSense 10,
humidification and
Standard
AirSense 10 and
SlimLine
AirSense 10,
humidification and
ClimateLineAir
cm H
2
O (hPa)
L/min
L/min
L/min
L/min
4
180
143
162
151
8
168
135
151
142
12
157
136
140
135
16
144
134
128
121
20
131
123
117
109
Symbols
The following symbols may appear on the product or packaging.
Read instructions before use.
Indicates a warning or caution.
Follow instructions before
use.
Manufacturer.
European Authorized Representative.
Batch code.
Catalog number.
Serial number.
Device number.
On / Off. Device weight.
Protected against finger sized objects and against dripping water when tilted up to 15 degrees
from specified orientation.
Direct current.
Type BF applied part.
Class II
equipment.
Humidity limitation.
Temperature limitation.
Non-ionising radiation.
China pollution control logo 1.
China pollution control logo 2.
Prescription only
(In the US, Federal law restricts these devices to sale by or on the order of a physician).