5
1.2
XCell™ ATF
System Control Scope and Objectives
The C410 v4 Controller provides the process control functionality of the stainless steel
XCell™ ATF
4 System,
XCell™ ATF
6 System,
XCell™ ATF 10 System and the XCell™ ATF 6
and 10 Single-
use devices. The XCell™ ATF
C410v4B controller is designed to:
1.
Control ATF flow rates in both pressure and exhaust cycles
2.
Provide a user interface capability for
XCell™ ATF
System control and monitoring
3.
Set-up and select operational parameters
4.
Display real-time process data and Alarms for error conditions
5.
Provide “
B
atch control” and user hierarchy
6.
Have PLC software upgrades in the field by replacing memory modules
7.
Have 3 major enclosures, an Electronics Box (E-Box) a Pneumatics Box (P-Box) and a
Power Box
8.
Allow the E-Box to operate an
XCell™ ATF
4,
6 and 10 systems and XCell™ ATF Single
-use
6 and 10 devices using custom software
9.
Allow the P-Box hardware and process parameters to be specific to the size of the
particular
XCell™ ATF
System being controlled.
Control principles are detailed in Section 6.
2.0
Utility Requirements
UTILITY
REQUIREMENT
ADDITIONAL CONSIDERATIONS
PRIMARY
COMPRESSED AIR
Maximum 90 psig /
6.1 bar
Oil free, dry, filtered gas, i.e., medical
grade air
SECONDARY
PRESSURE
Typically: 35 psi / 2.4bar,
Regulated oil free, dry, filtered air
VACUUM SERVICE
Minimum -12.5 psig
Vacuum supplied by a Repligen or
customer supplied local pump capable of
maintaining ~-12.5 psig with nominal
flow as follows:
XCell™ ATF4
- 40L/min
XCell™ ATF6 –
60L/min
XCell™ ATF10 –
200L/min
Pump should be clean room compatible
EXHAUST
Untreated discharge or user specified
ELECTRICAL
100-240V AC, 50/60Hz
NA
ENVIRONMENT:
Temperature: 0-50
°
C
Relative Humidity: 0
–
80
% RH
NA
STEAM (SIP)
~30lbs/hr.
Applicable only to a steamable
connection between
XCell™ ATF
and
bioreactor vessel
CONDENSATE DRAIN
For SIP condensate drainage