![REPLIGEN XCell ATF Series User Manual Download Page 32](http://html1.mh-extra.com/html/repligen/xcell-atf-series/xcell-atf-series_user-manual_1443425032.webp)
XCell
TM
ATF System with C410:V4B Controller
User Guide
28
The main screen displays the parameters in the following tables:
Field
Description
Date/Time
Displays current Date and Time
User ID
Displays current User ID
Entries in Main
Screen
Field entries by Administrator (or, if authorized, by Engineer) through the
Main screen: P-FLOW SP, E- FLOW SP, PV1, PRV1, PV2, PRV2
P-FLOW SP
Displays current P-Flow Set Point (SP) (LPM)
P-FLOW PV
When running, displays pump Flow rate Process Value (PV) of last pump
cycle (LPM)
P-FLOW
Calculated
Displays current P-Flow Calculated cycle time (Sec)
P-FLOW Actual
When running, displays Actual pump flow cycle Time of last pump cycle
(Sec)
E- FLOW SP
Displays current E-Flow Set Point (SP) (LPM)
E- FLOW PV
When running, displays pump exhaust Flow rate Process Value (PV) of last
pump cycle (LPM)
E-FLOW
Calculated
Displays current E-Flow Calculated cycle time (Sec)
E-FLOW Actual
When running, displays Actual pump exhaust cycle Time of last pump cycle
(Sec)
PV1
Displays current position Set Point of Flow Proportional Valve (0-100%)
PV2
Displays current position Set Point of Exhaust Proportional Valve (0-100%)
PRV1
Displays current setting of automatic pressure regulator (0 to 35psi)
PRV2
Displays current setting of automatic exhaust regulator (-15 to 0 psi)
P2
Displays current pressure between controller and Diaphragm Pump (PSI)
Controller
Status
Displays current controller status:
PUMP OFF
INFLATING
EXHAUSTING
PUMP WARNING
PUMP ALARM
SYSTEM STOP
Indicator
Description
PLC
Indicates Controller PLC is ON and in RUN mode
Pump ON
Indicates Diaphragm Pump is in RUN mode
Standby
Indicates Controller OFF/ON Status
SOL 1
Indicates Flow Direction Solenoid is OFF/ON
Bioreactor
Connected Bioreactor ID (input in setup basic)
Batch
Batch ID information (input in Batch Info)
Runtime
Time the current batch has been running (reset in Batch Info)
Current User
Displayed username of current login