23
Fitting the probe module and stylus
1. Push the Stop button to disable all machine motion. Make a habit of doing this whenever human
interaction is about to take place and manual mode is not activated.
2. Attach the probe module (SM25-2 or TM25-20) to the probe body, with the SM25-2 or TM25-20
text to the front.
Scanning - SM25-2
Touch Trigger - TM25-20
Scanning
3. Attach the 21 × 5 stylus to the probe module. The silver alignment marking should be towards you.
There is a click when the magnets lock the kinematic coupling in place.
Touch Trigger
3. Attach the TP20 module and stylus to the TM25-20, making sure that the alignment marks
correspond to the alignment marks on the TM25-20.
Scanning - SM25-2
Touch Trigger - TM25-20
4. Disengage the Stop button by twisting it clockwise