20
W.1: For further safety information, refer to the Equator Controller quick-start guide.
W.2: Be careful to connect the PCIexpress cable in the correct orientation, as failing to do so
can cause serious damage to the electronics. Refer to step 8.1.
W.3: Ensure all cables are connected prior to connecting to mains.
W.4: Power outputs must not exceed these limits
•
(P1) 24 V ± 3% 4 A
(MAX CONTINUOUS) 5 A (MAX PEAK)
•
(P2) 48 V ± 3% 4 A
(MAX CONTINUOUS) 8 A (MAX PEAK)
•
(P1 & P2) COMBINED (48 V & 24 V) CONTINUOUS POWER OUTPUT MUST BE LESS THAN 300 W
NOTE: Both ethernet ports (LAN 1, LAN 2) cannot be connected to the same network.
A-5504-0065
A-5504-0065
8.1
W.2
!
8.1
A-5504-0070
A-5504-0070
8.1
W.2
!
W.4
!
P1
P2
7.2