18
When the Equator has been removed from its transit packaging according to the instructions attached
to the box, please follow the instructions in the following sections to start the system.
Contents of boxes
The Equator gauging system is delivered in several boxes. The larger box contains the machine and
relevant accessories, other boxes contain the controller, stop button or joystick, probe kit and relevant
accessories.
Transporting the Equator
When placed in its original packaging, the Equator machine can be moved by forklift or sack truck.
Over shorter distances, it can be moved manually by two people using the hand holds on the side of
the larger box.
The system should be transported in the original Equator packaging to ensure that no parts get lost or
damaged.
The Equator machine should only ever be lifted by two people. Use the hand holds on the four corners
of the base casting.
Equator machine positioning
The Equator system needs to be placed on a flat and sturdy surface. It is advised that a 1 m diameter
clearance of other objects is established to avoid collisions.
It is suggested that the monitor and the joystick (if fitted) are located near the machine so that
operational information can be seen by the operator.
Equator controller positioning
The controller can be placed on either side of the Equator as preferred. It can be placed either
standing or lying down.
Ensure that the fan intakes at the front and outlets at the rear do not become blocked in any way. Allow
10 cm distance between fan intakes and outlets to any surface.
7.1 Unpacking