21
Pneumatic set-up
(continued)
Operation sequence (circuit 3)
Operation
sequence
Stylus
Solenoid
valve 1
Solenoid
valve 2
End of sequence
trigger
Monitor Variable to
set
Comment
1 Cover extend
Protected
0
0
Pressure switch
activation
Pressure
switch
Default state
2 “Air bleed” off
(extended)
Protected
0
0
Time
Pressure
switch
Sequence
hold time
3 “Air bleed” on
(extended)
Protected
0
1
Time
Pressure
switch
Bleed time System pressure
>
4.5 bar
(65.3 psi)
4 “Air bleed” off
(extended)
Protected
0
0
Time
Pressure
switch
Sequence
hold time
5 Cover retract
Accessible
1
0
Retract sensor
Retract
sensor
6 “Air bleed” on
(retracted)
Accessible
1
1
Time
Retract
sensor
Bleed time Allow control of
“air bleed” flow
7 “Air bleed” off
(retracted)
Accessible
1
0
Time
Retract
sensor
Sequence
hold time
1 Cover extend
Protected
0
0
Pressure switch
activation
Pressure
switch
“Default state”
stylus protected