Manual MWS 6
Additional Sensor
Sensors of any kind can be integrated as additional sensors. The information signal
must be applied as voltage in the range between 0 V and + 4 V or it has to be adapted
to MWS 6 by a special amplifier. Negative or higher voltages are blocked, but out of
security reasons, should be avoided in any case!
The sensor signals are linearised by software.
See Adjustment of an additional sensor (P.28).
Additional sensors can be voltage supplied by the weather station as long as its
current consumption is below 25 mA.
If current consumption exceeds this value, we cannot grant that MWS 6
weather station works properly !!!
If they need higher currents, the sensors must have an external power supply and
its ground has to be connected to the MWS 6-ground (Pin 1 of the connector socket).
You can call for adjustment instructions of the additional inputs at any time.
Additional Sensor
Note