11
15140-101 Rev. C
Setup
(continued)
Unpacking and Installation
(continued)
15. Install the Base Assembly onto the tracks of
the Table Top and slide the Guide Rail Covers
around the tracks. Refer to Figure SU-12.
16. Attach the Base Lamp Wire to the back of the
Power Supply Assembly. Refer to Figure SU-8.
17. Install the Microscope Assembly onto the top of
the Arm by sliding it into position, making sure
it is up against the stop. Then, tighten the Lock
Knob located on the right side of the Microscope
Assembly. Refer to Figure SU-13.
Note:
Do not adjust the microscope stop knob
behind the base of the microscope or the
vertex distance will cause misalignment of
focus and require re-calibration of the slit
lamp assembly.
Application of Input Power
WARNING:
CARE MUST BE TAKEN TO ARRANGE THE
CABLES FOR THE ACCESSORIES SUCH THAT THEY DO
NOT PRESENT A TRIPPING HAZARD TO THE EXAMINER
OR A DANGER TO THE PATIENT.
WARNING:
POSITION THIS INSTRUMENT SO THAT IT
IS NOT DIFFICULT TO OPERATE THE DISCONNECTION
DEVICE (PLUG).
1. After the unit is in its secure location, apply the
correct input voltage to the instrument using the
Power Cord.
Note:
The power inlet is located on the backside of
the Power Supply Assembly.
2. Activate ON/OFF Switch (I). Refer to Figure SU-14.
Note:
The ON/OFF Switch will illuminate green when
there is power to the unit. When the ON/OFF
Switch is set to off, the green light will turn off.
Disconnection of Input Power
1. At any time, the power switch can be set to OFF.
The unit does not have a power down sequence.
To terminate operation of this instrument, press
the ON / OFF switch to the OFF position (O).
2. If this instrument is intended to be OFF for an
extended period of time, it can be disconnected
from power by detaching the power cord from
the receptacle.
Figure SU-12 Install Base Assembly
Slide
Figure SU-13 Microscope Install
Turn Clockwise
Lock Knob
Microscope
Assembly
Slide
Figure SU-14 Power Supply Assembly
ON/OFF Switch
Illumination Level Control Knob