Fuse Replacement
Replace the fuses in the Power Input Module with
the fuses indicated in the Specifications section of
this manual.
1. Remove input power to the instrument.
2. Press down on the top tab in the middle of the
Power Input Module to release the Fuse Holder,
and gently pull out the Fuse Holder by gripping the
two small tabs. Refer to Figures MM-8 and MM-9.
3. Open the Door to the Fuse Holder by pulling it
down. Refer to Figure MM-9.
Note:
The Fuses will pop up when the door is
open, making removal easier.
4. Install new fuses into the Fuse Holder that is
indicated in the Specification section of this
manual.
5. Install the Fuse Holder by closing the door, and
pushing the Fuse Holder back until it snaps into
place.
Figure MM-8 Pull Out
Pull
Out
Tabs
Figure MM-9 Open Fuse Door
Open
Door
Fuses
Cleaning & Maintenance
(continued)