![Reichert XCEL 455 User Manual Download Page 20](http://html1.mh-extra.com/html/reichert/xcel-455/xcel-455_user-manual_1435042020.webp)
20
15140-101 Rev. C
Field of View
Magnification
Field of View (mm)
6.5 X
37.0
10 X
24.0
16X
15.0
25X
9.5
40X
6.0
Movement Ranges
Longitudinal (In/Out) 4.3 in. (110 mm )
Lateral (Left/Right)
4.3 in. (110 mm )
Vertical (Up/Down)
1.2 in. (30 mm )
Chinrest Range
2.8 in. (70 mm )
Table Dimensions
18.3 in. x 12.6 in. (465 mm x 320 mm)
Compliance
This slit lamp complies with:
IEC 60601-1
ISO 10939
ISO 15004-1
ISO 15004-2
Device Classification
Protection: Class I
Ingress Protection Rating: IXP0
Instrument Type: (IEC 60601-1) B
Operating Mode: Continuous
US FDA: Class II 21 CFR 886.1850
EU Medical Device Directive: Class I, Rule 12
CAN CMDR: Class I Rule 7, Clause 1
Disposal
This product does not generate any environmentally hazardous residues. At the end of its product life, follow
your local laws and ordinances regarding the proper disposal of this equipment.
Software Revision
There is no software installed in this unit.
Due to a policy of continuous development, we reserve the right to change specifications without notice.