Reichert XCEL  455 User Manual Download Page 7

7

15140-101 Rev. C

Introduction

Congratulations on the purchase of your new Xcel

®

 455 Slit Lamp from Reichert Technologies

®

.  

This user’s guide is designed as a training and reference manual for the operation and maintenance 

of the instrument.  We recommend that you read it carefully prior to use and follow the instructions to 

ensure optimum performance of your new instrument.  Properly trained eyecare professionals such as 

ophthalmologists, optometrists, opticians and eye care technicians should operate this instrument.  

Please retain this manual for future reference and to share with other users.  Additional copies can be 

obtained from your authorized Reichert

®

 distributor or from the Reichert customer service department at:

Tel: 716-686-4500

Fax: 716-686-4555

Email: [email protected]

Indications for use

The Xcel 455 Slit Lamp is an AC-powered slit lamp biomicroscope that is intended for use in examining the 

anterior eye segment, from the corneal epithelium to the posterior capsule.  It is used to aid in the diagnosis 

of diseases or trauma, which affect the structural properties of the anterior segment of the eye.

Contraindications

None.

Phototoxicity Information

Because prolonged intense light exposure can damage the retina, the use of the device for ocular examination 

should not be unnecessarily prolonged, and the brightness setting should not exceed what is needed to 

provide clear visualization of the target structures.  This device should be used with filters that eliminate UV 

radiation (< 400 nm) and, whenever possible, filters that eliminate short-wavelength blue light (< 420 nm).

The retinal exposure dose for a photochemical hazard is a product of the radiance and the exposure time. If 

the value of radiance were reduced in half, twice the time would be needed to reach the maximum exposure 

limit.

While no acute optical radiation hazards have been identified for slit lamps, it is recommended that the intensity 

of light directed into the patient’s eye be limited to the minimum level which is necessary for diagnosis.  Infants, 

aphakes and persons with diseased eyes will be at greater risk.  The risk may also be increased if the person 

being examined has had any exposure with the same instrument or any other ophthalmic instrument using a 

visible light source during the previous 24 hours.  This will apply particularly if the eye has been exposed to 

retinal photography.

Summary of Contents for XCEL 455

Page 1: ...XCEL 455 Slit Lamp User s Guide...

Page 2: ...nges in the product described in this manual without notice and without incorporating those changes in any products already sold ISO 9001 13485 Certified Reichert products are designed and manufacture...

Page 3: ...3 Filters 13 Slit Rotation 14 Illumination Inclination 14 Slit Centration 14 Cleaning Maintenance 15 Cleaning 15 External Cleaning 15 Forehead Chinrest Cleaning Disinfection 15 Cleaning the Glide Plat...

Page 4: ...ELECTRIC SHOCK THIS EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY MAINS WITH PROTECTIVE EARTH OR DAMAGE TO THIS INSTRUMENT AND OR INJURY TO THE OPERATOR OR PATIENT MAY OCCUR WARNING ENSURE THAT THE VOL...

Page 5: ...NG EMC AND NEEDS TO BE INSTALLED AND PUT INTO SERVICE ACCORDING TO THE EMC INFORMATION PROVIDED IN THE ACCOMPANYING DOCUMENTS CAUTION PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT CAN AFFECT MEDICAL...

Page 6: ...014 Date of Manufacture REF Catalog Number S N Serial Number Waste of Electrical and Electronic Equipment Compliance to Medical Device Directive 93 42 EEC 36FS Underwriter s Laboratories has independe...

Page 7: ...of the anterior segment of the eye Contraindications None Phototoxicity Information Because prolonged intense light exposure can damage the retina the use of the device for ocular examination should n...

Page 8: ...ial 13 Lamp Housing Cover 14 Fixation Light 15 Filter Lever 16 Slit Rotation Length Knob 17 Slit Rotation Scale 18 Microscope Lock Knob 19 Illumination Arm Lock Knob 20 Instrument Base Lock Knob 21 Gu...

Page 9: ...he accessories and store them in an appropriate place so that when they are needed they will be available Refer to Figure SU 4 8 Using the 3mm Hex Wrench connect the Ground Wire from the Power Supply...

Page 10: ...lumination Assembly and Arm Note The Illumination Assembly and Arm are con nected as one piece 14 Mount the Illumination Assembly and Arm onto the Base Assembly and secure it with the Screw using the...

Page 11: ...NGER TO THE PATIENT WARNING POSITION THIS INSTRUMENT SO THAT IT IS NOT DIFFICULT TO OPERATE THE DISCONNECTION DEVICE PLUG 1 After the unit is in its secure location apply the correct input voltage to...

Page 12: ...Chin Rest Assembly until the patient s canthus is in line with the Canthus Mark on the chin rest post Refer to Figures IN 2 and IN 3 7 Microscope elevation is adjusted by rotating the joystick and obs...

Page 13: ...slit length is adjusted by rotating the Slit Length Rotation Dial The dial has five stops for adjustments They are 12 9 5 3 1 and 0 3 mm diameter and continuous length Refer to Figure IN 8 Filters Th...

Page 14: ...mination Assembly by depressing the Inclination Latch Release and pulling the base of the Illumination Assembly toward the operator Refer to Figure IN 10 Slit Centration When the centering screw is lo...

Page 15: ...d rest with an alcohol wipe and change the chin rest papers after each patient Cleaning the Glide Plate If the Glide Plate is dirty it may cause a rough feeling when maneuvering the base of the slit l...

Page 16: ...Holder and Lamp Refer to Figure MM 5 5 Grip the Lamp by the Metal Disk that is attached to it and pull it out of the Lamp Holder Refer to Figure MM 6 6 Replace the Lamp with a new one and install it...

Page 17: ...gripping the two small tabs Refer to Figures MM 8 and MM 9 3 Open the Door to the Fuse Holder by pulling it down Refer to Figure MM 9 Note The Fuses will pop up when the door is open making removal e...

Page 18: ...ation Light Harness not plugged into the Power Supply Assembly Ensure the Fixation Light Harness is properly seated in the Power Supply Assembly Defective Power Supply Replace the Power Supply Assembl...

Page 19: ...35 C 95 F Relative Humidity 30 to 75 Atmospheric Pressure 80 kPa 23 6 in Hg to 106 kPa 31 3 in Hg Transportation Storage Temperature 20 C 4 F to 70 C 158 F Relative Humidity 10 to 80 Atmospheric Pres...

Page 20: ...tion Protection Class I Ingress Protection Rating IXP0 Instrument Type IEC 60601 1 B Operating Mode Continuous US FDA Class II 21 CFR 886 1850 EU Medical Device Directive Class I Rule 12 CAN CMDR Clas...

Page 21: ...n such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 Class A The Xcel 455 uses RF energy only for its internal function Therefore its RF e...

Page 22: ...ity should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1kV Differential 2kV Common 1kV Differential 2kV Common Mains power quality should be that of a typical commercia...

Page 23: ...distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur...

Page 24: ...m output power of the communications equipment Max Output Pow er of Transmitter W Separation m 150kHz to 80 MHz d 3 5 V1 Sqrt P Separation m 80 to 800 MHz d 3 5 E1 Sqrt P Separation m 800MHz to 2 5GHz...

Page 25: ...RRANTY If notified promptly in writing of any action brought against the purchaser based on a claim that the instrument infringes a U S Patent Reichert will defend such action at its expense and will...

Page 26: ...26 15140 101 Rev C Notes...

Page 27: ...27 15140 101 Rev C Notes...

Page 28: ...716 686 4545 Email reichert information ametek com www reichert com AMETEK GmbH Business Unit Reichert Carl von Linde Strasse 42 85716 Unterschleissheim Munich Germany Email info reichert de ametek c...

Reviews: