26
8. Program overview
Stimulation
Frequency
(Hz)
Pulse
duration
(μS)
Extra information
P1
Conventional
TENS
110
80
P2
Low frequency
TENS
2
250
P3
Combination
conventional
and low
frequency
TENS
110
80
left channel (P1)
2
250
right channel (P2)
P4
HAN
100
150
phase 1 / 3 seconds
2
200
phase 2 / 3 seconds
P5
HAN combina-
tion
100/2
150/200
left channel (P4)
110
80
right channel (P1)
P6
Burst
100
180
8 pulses, 2x per sec.
P7
Pulsewidth
modulation
110
60 tot
100
1 sec.
from min. to max.
P8
Frequency
modulation
2 tot 100
80
5 sec.
from min. to max.
P9
Depth TENS
3.636k
75
4 pulses,
100x per sec.
P10
Muscle
stimuation
100
150
ramp up: 1 sec.
100
150
stimulation: 4 sec.
100
150
ramp down: 1 sec.
pause: 3 sec.
Summary of Contents for i-Pulz