If there is no or limited connectivity between the device
and your skin, you are able to set up the intensity to a
maximum of 7.5mA. If you try to go beyond the 7.5mA,
the device will produce a beeping sound and the inten-
sity will be reduced to 00,0 mA. You will also see a war-
ning sign on the display. This problem is caused by the cables or elec-
trodes. Check that the cables and electrodes are properly connected
and if necessary replace them.
If you are only using 1 channel, e.g. the left channel, you should
use the left Up / Down button.
If you have finished setting up the intensity, the automatic key lock
will activate after 15 seconds. The keypad will be activated again by
pressing the Down-button once.
An overview of all available programs can be found in the back of the
manual at the technical specifications.
3.4 Treatment period
Regarding the treatment period you must strictly adhere to the
instructions from your doctor or specialist. If you need a longer treatment
period, you should discuss this with your doctor or specialist before
changing your stimulation patterns.
Close the flip of the device during treatment to prevent any damage to
the display.
3.5 Stopping your daily treatment
If you have activated the timer, the i-Pulz stimulator will switch off
automatically after the treatment has run its course. You can also stop
the program manually by pressing the MODE button or by tuning down
the stimulation until it reaches 00,0mA. When the current reaches
00,0mA the device will switch off after a few minutes in order to pre-
serve the batteries. The device can also be switched off by pressing the
On / Off button.