29
9.3 Immunity Guidance for Users
Guidance and Manufacturer’s declaration-electromagnetic immunity
The i-Pulz (ES-2) is intended for use in the electromagnetic environment specified
below. The customer or the user of i-Pulz (ES-2) should assure that it is used in the
following environment:
Immunity Test
IEC 60601-1-2
Test Level
Compliance
Level
Electromagnetic
Environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV con-
tact
±8 kV air
Floors should be made of
wood, concrete, or ceramic
tile. If floors are covered
with synthetic material, the
relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
Not
applicable
The device is battery
powered
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Not
applicable
The device is battery
powered
Voltage dips,
short interrupti-
ons, and voltage
variations on po-
wer supply input
lines
IEC 61000-4-11
<5% Un
(>95% drop in Un)
for 0,5 cycle
40% Un
(60% drop in Un)
for 5 cycles
70% Un
(30% drop in Un)
for 25 cycles
<5% Un
(>95% drop in Un)
for 5 seconds
Not
applicable
The device is battery
powered
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commer-
cial or hospital environment.
Summary of Contents for i-Pulz
Page 1: ...i Pulz User Manual i Pulz User Manual...
Page 2: ......