14
3.2 Placing of the electrodes
Keep to the instructions of your doctor or specialist when placing
the electrodes.
Follow the instructions on the packaging of the electrodes or on its user
manual.
3.3 Setting up the stimulator
After the electrodes are placed properly, you can activate the device. The
device starts up with the program you used last. During the trial period
you and your doctor/specialist have decided which program(s) is best
suited for you.
In the back of this manual you will find a summary of the standard pro-
grams.
Every time you press the MODE button the next program will appear (it
is an ongoing list, you will reach your starting position eventually).
Check if the timer is set properly. By default, the timer is set to 30
minutes per program, unless your doctor or specialist has altered this.
The timer you set for a program will be saved and used automatically
when you select that program again.
After you have selected the desired program, you can set up its inten-
sity. Pressing the Up-button briefly will increase the intensity in small
steps of 0,5mA. When you keep the up-button pressed, the intensity
will increase faster.
Caution: for safety regulations it is not possible to increase the in-
tensity beyond 20mA when keeping the Up-button pressed. Only
when you release the button briefly you can increase the intensity
again, but only 10mA per time.