The above note also applies to testing at 1000V. In addition the following
applies.
Note:
Make sure that the circuit under does not include components
Which will be damaged by the 1000V applied. Many normal components
of an installation are likely to be damaged if tested at 1000V. Examples are
power factor correction capacitors, low voltage mineral insulated cables,
electronic light dimmers, electronic ballasts and starters for fluorescent
lamps etc…
Lock Power on Feature
For hands free operation a lock power on feature is incorporated in the
press to test button. Set the LOCK button to lock test voltage, pressing it
again will turn the tester off.
Low Resistance (Continuity) Measurements
1. Set the range switch to 200Ω •))) Position
2. Connect the red test lead to the V Ω terminal and back to the
.
COM terminal.
3. Connect the tips of the test leads to both ends of the circuit under test.
Read resistance in Ω on the LCD.
4. When the impedance on circuit is below approximately 40Ω.
.
It will indicate by a continuous beeper.
AC/DC Voltage Measurements
1. Set the range switch to ACV or DCV position
2. Connect red test lead to “V Ω” terminal and black test lead
.
to terminal “COM”.
3. Connect test prods of test leads IN PARALLEL to the circuit
.
being measured.
4. Read the voltage value on LCD.