18
Data Stream Output
Using Terminal application, a data stream can be viewed from the RS-232
output, connect RS-232/USB cable between the product and terminal/PC
and use the following setting to view the data stream.
Terminal Settings:
• Bits per second: 9600
• Data bits: 8
• Parity: None
• Stop bits: 1
41110000000335 = Upper Value, 335 FPM
Reading
Decimal Place 0-3
Positive (+) = 0,
Negative (-) = 1
°C= 01, °F = 02, m/s= 08, Knot= 09,
Km/h= 10, FPM= 11, mph= 12
1 = Upper display value,
2 = Lower display value
4
41010100000221 = Canal T1, +22.1°C
Lecture
Position décimale 0-3
Positif (+) = 0, Negatif (-) = 1
°C = 01, °F = 02
Canal T1 = 1, T2 = 2, T3 = 3, T4 = 4
4
3.5mm Terminal to RS-232 Serial Connection
RS-232
Mono Jack
Pin 2 (RX)
Pin 4 (DTX)
Sleeve
Tip
REED Instruments
1-877-849-2127 | [email protected] | www.reedinstruments.com