10
11
TC input:
Cold junction compensation (CJC)
via internally mounted sensor ................. < ±1,0 °C
Sensor error detection, all TC types ........... Yes
Sensor error current:
when detecting ....................................... Nom. 2 µA
else
..........................................................
0
µA
Current input:
Measurement range .................................... -1...25 mA
Programmable measurement ranges .......... 0...20 and 4...20 mA
Input resistance ........................................... Nom. 20
Ω
+ PTC 50
Ω
Sensor error detection:
Loop break 4...20 mA ............................. Yes
Voltage input:
Measurement range .................................... -20 mV...12 VDC
Programmable measurement ranges .......... 0...1 / 0.2...1 / 0...5 / 1...5 /
0...10 and 2...10 VDC
Input resistance ........................................... Nom. 10 M
Ω
Current output:
Signal range (span) ...................................... 0...20 mA
Programmable signal ranges ...................... 0...20 / 4...20 /
20...0 and 20...4 mA
Load (max.) .................................................. 20 mA / 800
Ω
/ 16 VDC
Load stability ...............................................
≤
0.01% of span / 100
Ω
Sensor error detection ................................ 0 / 3.5 / 23 mA / none
NAMUR NE 43 Upscale / Downscale ......... 23 mA / 3.5 mA
11
Auxiliary supplies:
2-wire supply (terminal 44...43) ................... 25...16 VDC / 0...20 mA
Max. wire size .............................................. 1 x 2.5 mm
2
stranded wire
Screw terminal torque ................................. 0.5 Nm
Relative humidity ......................................... < 95% RH (non-cond.)
Dimensions, without display front (HxBxD) .. 109 x 23.5 x 104 mm
Dimensions, with display front (HxBxD) ...... 109 x 23.5 x 116 mm
Protection degree (enclosure / terminals) ... IP50 / IP20
Weight ......................................................... 170 g / 185 g with PGMMOD00
RTD, linear resistance and potentiometer input:
Input for RTD types:
Pt10, Pt20, Pt50, Pt100, Pt200, PT250, Pt300, Pt400, Pt500, Pt1000
Ni50, Ni100, Ni120, Ni1000
Cable resistance per wire (max.), RTD ........ 50
Ω
Sensor current, RTD .................................... Nom. 0.2 mA
Effect of sensor cable resistance
(3- / 4-wire), RTD ......................................... < 0.002
Ω
/
Ω
Sensor error detection, RTD ....................... Yes
Short circuit detection, RTD ........................ < 15
Ω
10
EMC immunity influence ..................................... < ±0.5% of span
Extended EMC immunity:
NAMUR NE 21, A criterion, burst ....................... < ±1% of span
Input
type
Min.
value
Max.
value
Standard
Pt100
Ni100
Lin. R
Potentiometer
-200°C
-60°C
0
Ω
10
Ω
+850°C
+250°C
10000
Ω
100 k
Ω
IEC60751
DIN 43760
-
-
Type
Min.
value
Max.
value
Standard
B
E
J
K
L
N
R
S
T
U
W3
W5
LR
+400°C
-100°C
-100°C
-180°C
-200°C
-180°C
-50°C
-50°C
-200°C
-200°C
0°C
0°C
-200°C
+1820°C
+1000°C
+1200°C
+1372°C
+900°C
+1300°C
+1760°C
+1760°C
+400°C
+600°C
+2300°C
+2300°C
+800°C
IEC 60584-1
IEC 60584-1
IEC 60584-1
IEC 60584-1
DIN 43710
IEC 60584-1
IEC 60584-1
IEC 60584-1
IEC 60584-1
DIN 43710
ASTM E988-90
ASTM E988-90
GOST 3044-84