INSTALLATION & REMOVAL OF THE RELAY BOARD
To install the relay board, locate the relay opening at the lower right-hand
corner, on the back of the Gemini. Pull the tab down while sliding the board into
the two slots in the housing. The relay board will seat into the unit, allowing the
tab to return to its original position. To remove the relay board, pull down on the
tab just enough to allow the relay board to slide out. Grasp the terminal connector
and slowly pull until the relay board is removed.
NOTES:
1. SENSOR VOLTAGE AND CURRENT
The +12 V sensor supply voltage on the
“DC OUT”
terminal is nominal with
+/-25% variation due to line and internal load variations. All RLC sensors will
accommodate this variation.
2. HI/LO FRQ SELECTION
The HI/LO FRQ selection switch must be set on
“LO FRQ”
when switch
contacts are used to generate count input signals. Since the
“LO FRQ”
mode
also provides very high immunity against electrical noise pickup, it is
recommended that this mode also be used, whenever possible, with electronic
sensor outputs. The
“LO FRQ”
mode can be used with any type of sensor
output, provided pulse widths never decrease below 5 msec, and the count rate
does not exceed 100 cps.
3. V
IL
and V
IH
levels given are nominal /-10% when the voltage on “DC
OUT” terminal is +12 VDC. These nominal values will vary in proportion to
the variations in the “DC OUT” terminal voltage, which are caused by line
voltage and load changes.
4. When shielded cable is used, the shield should be connected to
“COMM.”
at
the counter and left unconnected at the sensor end.
5. Input B set-up is identical to that of Input A. Input B is for Logic Inputs only.
6. Inputs A and B can accept source pulses from other circuits up to +28 V in
amplitude. For voltages above +28 V, a limiting resistor and zener diode
should be used to limit the voltage at the input.
–36–
Summary of Contents for GEMINI 1000
Page 33: ...GEMINI 1000 BLOCK DIAGRAM...
Page 34: ...GEMINI 2000 BLOCK DIAGRAM...
Page 39: ...GEMINI 1000 CONNECTIONS CONFIGURATION SWITCH SET UPS FOR VARIOUS SENSOR OUTPUTS SEE NOTE 5...
Page 40: ...GEMINI 2000 CONNECTIONS CONFIGURATION SWITCH SET UPS FOR VARIOUS SENSOR OUTPUTS SEE NOTE 5...
Page 53: ...APPENDIX F GEMINI 1000 PROGRAMMING CHART CARD...
Page 54: ...APPENDIX F GEMINI 1000 PROGRAMMING CHART CARD...
Page 55: ...APPENDIX F GEMINI 2000 PROGRAMMING CHART CARD...
Page 56: ...APPENDIX F GEMINI 2000 PROGRAMMING CHART CARD...