The “LS.” Input allows the CUB7P to operate at speeds up to 10 KHz when

driven by bi-polar outputs or external circuits having an output impedance of
3.3 K

or less. Input drive voltage must be limited to 3 V maximum to avoid

damage to the counter. CMOS and TTL Logic outputs can be loaded with a
resistor (R

L

) to limit drive voltage, or a voltage divider can be used as shown

for the PNP O.C. Transistor output.

L.S. INPUT; LOGIC VERSIONS (10 KHz 

MAX

.)

TTL OR CMOS

PNP O.C. TRANSISTOR

NPN O.C.

OUTPUT

OR BI-POLAR OUTPUT

TRANSISTOR

FIG 1

FIG 2

FIG 3

R values for

Fig 2 & 3

+V

R

+5 V

2.2 K

+12 V 10 K
+18 V 16 K
+24 V 24 K

INSTALLATION ENVIRONMENT

The unit should be installed in a location that does not exceed the maximum

operating temperature and provides good air circulation. Placing the unit near
devices that generate excessive heat should be avoided.

The bezel should be cleaned only with a soft cloth and neutral soap product.

Do NOT use solvents.

Continuous exposure to direct sunlight may accelerate the aging process of

the bezel.

Do not use tools of any kind (screwdrivers, pens, pencils, etc.) to operate

the push buttons of the unit.

INSTALLATION

The CUB7P meets NEMA 4X/IP65 requirements for indoor use when

properly installed. The units are intended to be mounted into an enclosed
panel. The viewing window and reset button are factory sealed for a
washdown environment. A sponge rubber gasket and mounting clip are
provided for installing the unit in the panel cut-out.

The following procedure assures proper installation:

1. Cut panel opening to specified dimensions. Remove burrs and clean around

panel opening.

2. Carefully remove and discard the center section of the gasket.
3. Slide the panel gasket over the rear of the counter body to the back of the

bezel. Install CUB7P unit through the panel cut-out.

4. Insert the mounting screws onto both sides of mounting clip. Tip of screw

should NOT project from hole in mounting clip.

5. Slide the mounting clip over the rear of the unit until the clip is against the

back of the panel. The mounting clip has latching features which engage
into mating features on the CUB7P housing. 

6. Note: It is necessary to hold the unit in place when sliding mounting clip

into position.

7. Alternately  tighten  each  screw  to  ensure  uniform  gasket  pressure.

Visually inspect the front panel gasket. The gasket should be compressed to
about 75 to 80% of its original thickness. If not, gradually turn mounting
screws to further compress gasket.

8. If the gasket is not adequately compressed and the mounting screws can no

longer be turned, loosen mounting screws, and check that the mounting clip
is latched as close as possible to the panel.

9. Repeat from step #5 for tightening mounting screws.

Summary of Contents for CUB7P

Page 1: ...s that appear in the manual or on equipment must be observed to ensure personal safety and to prevent damage to either the instrument or equipment connected to it If equipment is used in a manner not...

Page 2: ...erature to 40 C Must use NEC Class 2 or SELV rated power supply 4 SIGNAL INPUT LS terminal 4 See Count Edge under Programming for incrementing edge of input signal Contact Input CUB7P0xx 30 Hz max fro...

Page 3: ...uch as ferrite suppression cores is effective Install them on Signal and Control cables as close to the unit as possible Loop the cable through the core several times or use multiple cores on each cab...

Page 4: ...when properly installed The units are intended to be mounted into an enclosed panel The viewing window and reset button are factory sealed for a washdown environment A sponge rubber gasket and mounti...

Page 5: ...max voltage drop 0 5 V when ON The OFF state leakage current must be less than 2 A Note The RC protection circuit on the RST Input causes a delay of approximately 15 msec in Reset response L S INPUT C...

Page 6: ...t versions of the CUB7P require an external 9 to 26 VDC power supply The external supply is connected between the V and common terminals as shown in the drawing APPLICATION TOTAL YARDS OF MATERIAL A f...

Page 7: ...For Sales and Support Contact Walker EMD Toll free 800 876 4444 Tel 203 426 7700 Fax 203 426 7800 www walkeremd com 7 This page intentionally left blank...

Page 8: ...or products containing components manufactured by RLC and based upon personal injuries deaths property damage lost profits and other matters which Buyer its employees or sub contractors are or may be...

Reviews: