6.2  MODULE 2 - S

ECONDARY

 F

UNCTION

 P

ARAMETERS

 (

2-SEC

)

SEL

Max Capture 

Delay Time

Max Display 

Enable

Min Display 

Enable

Access Code 

For Service 

Operations

Min Capture 

Delay TIme

Factory 
Service 

Operations

2-SEC

HI-En

HI-t

LO-En

LO-t

FCS

CodE

Pro

PARAMETER MENU

MIN DISPLAY ENABLE

YES

NO

2.0

HI-t

NO

LO-En

NO

FCS

MAX CAPTURE DELAY TIME

When the Input Display is above the present MAX value for the entered 

delay time, the meter will capture that display value as the new MAX reading. 
A delay time helps to avoid false captures of sudden short spikes.

2.0

LO-t

MIN CAPTURE DELAY TIME

When the Input Display is below the present MIN value for the entered delay  

time, the meter will capture that display value as the new MIN reading. A delay 
time helps to avoid false captures of sudden short spikes.

0.0

   

to

  

999.9

 

seconds

NO

HI-En

MAX DISPLAY ENABLE

0.0

   

to

  

999.9

 

seconds

Select 

YES

 to perform any of the Factory Service Operations shown below.

FACTORY SERVICE OPERATIONS

yES

NO

YES

NO

Enables the Maximum Display Capture capability.

Enables the Minimum Display Capture capability.

The CUB5RT uses stored resistance calibration values 

to provide accurate temperature measurements. Over 
time, the electrical characteristics of the components 
inside the meter could slowly change. The result is that 
the stored calibration values may no longer accurately 

define the input circuit. For most applications, recalibration every 1 to 2 years 
should be sufficient.

Calibration of the CUB5RT involves a resistance calibration. Allow 30 

minute warm up before performing any calibration related procedure. The 
following procedures should be performed at an ambient temperature of 15 to 
35 °C (59 to 95 °F).

Calibration should only be performed by individuals experienced in 

calibrating electronic equipment.

CAUTION: The accuracy of the calibration equipment will directly affect the 

accuracy of the CUB5RT.

10 OHM RTD Range Calibration

1. Set the Input Range Jumper to 10 ohm.
2. With the display at 

CodE 48

, press and hold the 

SEL

 button for 2 seconds. Unit 

will display 

CAL NO

.

3. Press the 

RST

 button. Display reads 

CAL r10

.

4. Press the 

SEL

 button. Display reads 

00r

.

5. Apply a direct short to terminals INP+, EXC, and COMM using a three wire 

link. Press 

SEL

. Display reads 

CALC

 for about 15 seconds.

6. When the display reads 

15.0r

, apply a precision resistance of 15 ohms (with an 

accuracy of 0.01% or better) to terminals INP+, EXC, and COMM using a 
three wire link. Press 

SEL

. Display reads 

CALC

 for about 15 seconds.

7. When display reads 

CAL NO

, press the 

SEL

 button to exit calibration, or 

proceed to the 100 ohm RTD Range Claibration.

100 OHM RTD Range Calibration

1. Set the Input Range Jumper to 100 ohm.
2. With the display at 

CodE 48

, press and hold the 

SEL

 button for 2 seconds. Unit 

will display 

CAL NO

.

3. Press the 

RST

 button until the display reads 

CAL r100

.

4. Press the 

SEL

 button. Display reads 

0.0r

.

5. Apply a direct short to terminals INP+, EXC, and COMM using a three wire 

link. Press 

SEL

. Display reads 

CALC

 for about 15 seconds.

6. When the display reads 

300.0r

, apply a precision resistance of 300 ohms (with 

an accuracy of 0.01% or better) to terminals INP+, EXC, and COMM using 
a three wire link. Press 

SEL

. Display reads 

CALC

 for about 15 seconds.

7. When display reads 

CAL NO

, press the 

SEL

 button to exit calibration.

CALIBRATION

48

CodE

RESTORE FACTORY DEFAULT SETTINGS

Entering Code 66 will overwrite all user settings with 

the factory settings. The meter will display 

rESEt

 and then 

return to 

CodE 00

. Press 

SEL

 button to exit the module. 

Pressing both the 

SEL 

and the 

RST 

button on power-up 

will also load the factory settings and display 

rESEt

. This allows operation in 

the event of a memory failure or corruted data.

66

CodE

RESISTANCE DISPLAY MODE

Entering Code 85 will place the CUB5RT in a resistance 

display mode. This mode is useful for diagnostic purposes 
before and after calibration, or to display the measured 
resistance of a connected RTD probe. If the RTD type is set 

for 

Cu427

 with the jumper set to the 10 ohm position, the display will read 

resistance in 

0.000

 ohms resolution. For all other RTD types, with the jumper in 

the 100 ohm position, the display will read in 

0.00

 ohms resolution.

Re-entering code 85 toggles the display back to the temperature display mode 

without having to remove power from the meter. If power is removed, the 
display always returns to the temperature display mode when power is reapplied.

85

CodE

Summary of Contents for CUB5RT

Page 1: ...this literature or on equipment must be observed to ensure personal safety and to prevent damage to either the instrument or equipment connected to it If equipment is used in a manner not specified by...

Page 2: ...es for additional information 8 MEMORY Nonvolatile E2PROM memory retains all programming parameters and max min values when power is removed 9 CONNECTIONS Wire clamping screw terminals Wire Strip Leng...

Page 3: ...Insert the unit into the panel cutout While holding the unit in place push the panel latch over the rear of the unit so that the tabs of the panel latch engage in the slots on the case The panel latch...

Page 4: ...MI suppression devices or equivalent are recommended Fair Rite part number 0443167251 RLC part number FCOR0000 Line Filters for input power cables Schaffner FN2010 1 07 Red Lion Controls LFIL0000 6 To...

Page 5: ...NNECTIONS OSNK1 OSNK2 COM 3 WIRE RTD COMM INP EXC 2 WIRE RTD COMM INP EXC JUMPER CAUTION Power input common and sensor input common are NOT isolated from user input common In order to preserve the saf...

Page 6: ...ges to memory and return the meter to the Display Mode If power loss occurs before returning to the Display Mode verify recent parameter changes PROGRAMMING TIPS It is recommended to start with Module...

Page 7: ...be automatically loaded for the selected type The position of the Input Range Jumper must match the RTD type selected F SCALE TEMPERATURE SCALE Select the temperature scale This selection applies for...

Page 8: ...ess the SEL button Display reads 00r 5 Apply a direct short to terminals INP EXC and COMM using a three wire link Press SEL Display reads CALC for about 15 seconds 6 When the display reads 15 0r apply...

Page 9: ...rogramming Immediate Access Entering Code 222 allows access regardless of security code 6 3 MODULE 3 DISPLAY AND FRONT PANEL BUTTON PARAMETERS 3 dSP SEL Pro Front Panel Display Select Enable Display U...

Page 10: ...Balanced Hys LO bL OUTPUT STATE OFF ON Hys SP SP Hys OFF TRIGGER POINTS High Acting Unbalanced Hys HI Ub 2 HYS n HYSTERESIS VALUE 1 to 59999 Enter desired hysteresis value See Setpoint Output Figures...

Page 11: ...ns Meter Address Abbreviated Printing 5 SEr bAUd dAtA PAr Addr Abbr OPt Pro The Serial Setup Parameters are only active when one of the optional serial communications programming cards is installed in...

Page 12: ...it Only Programming Security Code CodE Display Decimal Point RTD Type tYPE 1 INP Temperature Scale SCALE dECPt USrIN User Input Function Filter Setting Display Offset Value OFSEt FILtr Filter Band bAN...

Reviews: