6
7
Install the AED Sentinel Monitoring Hardware
Install the battery pack
1.
Insert batteries.
Make sure the battery pack is turned off. Slide the cover to remove
it and insert all the included batteries, taking care to position them as shown in the
battery compartment. Replace the cover.
2.
Attach the battery pack to the cabinet.
Magnetically attach the battery pack to the
door hinge side of the cabinet, midway between the top and bottom, with the power
switch and power cable facing up.
Install the communications module
1.
Connect the ribbon cable to the
communications module.
Find the
ribbon cable. Each end has a black
plastic connector. Find the connector
labeled END 2 and firmly connect
it into the rectangular port of the
communications module. The cable
will feed out toward the magnet side
of the communications module.
2.
Connect the power cable.
Plug the
battery pack power cable into the
communications module using the
round port next to the ribbon cable connector. Adjust the position of the battery pack,
if necessary.
3.
Connect the antenna cable to the communications module.
Screw the free end
L-connector of the antenna cable onto the threaded receptor on the communications
module. Make sure the connector aligns with the receptor so the prong in the middle
of the cable doesn’t bend. Point the antenna cable L-connector toward the inside of
the cabinet and hand-tighten it.
4.
Attach the communications module to the cabinet.
Magnetically attach the
communications module toward the top of the cabinet wall on the hinged side above
the battery pack. Make sure the round button and LED light are facing outward,
toward the cabinet door. Push the communications module toward the back wall of
the cabinet so you can see its LED light with the cabinet door closed.