Install the AED Sentinel Monitoring Hardware
2.
Check the camera lens and AED position.
Close the cabinet door and, through the
glass, view the three aiming notches on the back of the camera to check the position
of the camera lens in relation to the AED’s status indicator.
3.
Aim horizontally.
Viewing the status indicator at eye level, use the aiming notch at the
bottom of the camera to check whether the camera lens points at the center of the
status indicator. If it doesn’t, open the door and, using the left and right sides of the
top mounting block (NOT the camera or rails), adjust the mounting system along the
cabinet door or move the AED left or right. Close the door and check the position
again. Repeat until the aiming notch aligns horizontally with the center of the
status indicator.
AIMING
NOTCH
STATUS
INDICATOR