Activate the AED Sentinel Monitoring Unit
2.
Turn on the battery pack:
Turn on the battery pack by moving the switch next to the
power cable to ON. This begins the 3-minute AED Sentinel activation sequence. The
status light on the communications module will turn solid green showing the AED
Sentinel monitoring unit is ready to activate.
3.
Activate the AED Sentinel monitoring unit:
Press the button on the communications
module and close the cabinet door, being careful not to pinch any cables.
Look at the status light on the communications module which will go through the
following sequence:
Alternating red and green blink:
The AED Sentinel monitoring unit is capturing and
sending an image. This will take about 30-60 seconds.
Then either …
Blinking green:
The AED Sentinel monitoring unit has captured an image and
successfully transmitted it to AED Sentinel Manager. The light will continue to blink for
about 1 minute.
OR
Blinking red:
The AED Sentinel monitoring unit either couldn’t capture an image or
couldn’t transmit it. See Troubleshooting below. The light will continue to blink for
about 1 minute.
Then …
Solid green:
The AED Sentinel monitoring unit is completing its activation sequence.
The solid green light will stay on for the remainder of the 3-minute AED Sentinel
activation period.
COMMUNICATIONS
MODULE STATUS
INDICATOR
BATTERY PACK
ON/OFF SWITCH