21 / 25
C Reset
Resetting the controller to factory default settings
can be done from this setting.
Menu Button Insert Password Downward
Button Others Enter Button
Downward Button Reset
Controller will set to factory default settings in 3
secs. and return to main Screen.
Factory settings will clear following settings to default:
1- Loop current values to default
2- Clock settings to default
3- Calibration values to default
4- Changed password to default
5- MTC to 25
C
6- Display settings to default
7- Relays output settings to default
D Logger Data
Controller comes
with built in feature of logging
conductivity or TDS values with respect to date, time
and corresponding temperature.
Menu Button Insert Password Downward
Button Others Enter Button
Upward Button Logger Data
Logger data can view up to 60 different values for
conductivity or TDS. This will save the data which
user requires at the corresponding time, date and
temperature and review it from Logger data when
needed. In logger data, values with (u) units
indicates Microsiemens /cm, values with (m) units
indicates Millisiemens /cm , values with (p) units
indicate Parts / million, values with (t) units indicates
Parts / thousand.
These logged values can be deleted by pressing
menu button for
5 secs. And controller will clear all
its logged data and indicates that values are cleared.
Note: Pressing enter button on main Screen will save
the values in logger data and indicates that value is
stored.