RCB KE-CD 1000 Product Instruction Manual Download Page 21

 

16 / 25 
 

B.1        Status (MTC / ATC) 

 

Controller can be compensated both  manually and 
automatically  according  to  conductivity  sensor. 
Automatic Temperature Compensation (ATC) needs 
to  be  selected  if  the  conductivity  sensor  has  RTD 
PT1000  feature.  CDS  1000  series  contains  RTD 
PT1000  feature  for  ATC.  If  the  conductivity  sensor 

doesn’t have the RTD PT1000 feature than MTC will 

be selected automatically.  

Menu Button         Insert Password         Downward Button       Calibration        Enter  
Button       Downward Button        Temperature         Enter Button         Downward  
Button        Status (MTC/ATC)         Enter Button 

Press enter to change the status to 

MTC / ATC

.   

 

Note:

 

Don’t forget to enter 

temperature coefficient value. (See below option: b.2)   

 

 
 

B.2       Temp. Setting 

 

Manual  Temperature  Compensation  (MTC)  can  be 
performed using this feature.  

Menu Button         Insert Password         Downward 
Button         Calibration        Enter Button        
Downward Button        Temperature          
Enter Button         Downward Button         
Temperature Setting         Enter Button 

Temperature  setting  has  two  options  as  set 
temperature and temperature coefficient. 

 

b.1       Set Temperature 

 

Temperature  value  required  for  MTC  can  be  provided  in 
this setting.  

Menu Button          Insert Password         Downward  
Button         Calibration        Enter Button      Downward 
Button        Temperature         Enter Button         Downward 
Button        Temperature Setting         Enter Button  
Downward Button        Set Temperature         Enter Button 

Press enter for MTC temperature value Box. Upward   and  
downward  button  will  increase  and  decrease  the  main 
value  by  1.  Forward  and  menu  button  will  increase  and 
decrease decimal (.0) value by 1. For temperatures larger 
than 100 it is recommended to reverse from end value by 
pressing downward button. Press enter to save value and 
escape for main display screen.      

 

 

Summary of Contents for KE-CD 1000

Page 1: ...Conductivity TDS Controller KE CD 1000 PIM R0 2020 RCB Chemicals Services All rights Reserved PRODUCT INSTRUCTION MANUAL...

Page 2: ...lacement parts specified by RCB Chemicals representative or call at provided number for assistance Unauthorized parts and procedures can affect the product s performance and place the safe operation o...

Page 3: ...Section 4 Controller Electrical Connections 4 1 General Guidelines 4 4 2 Power 4 4 3 Sensor Electrode 5 4 4 PLC Autologger Connection 6 4 5 Relay HI LO Connection 6 Section 5 Keypad Function 5 1 Gene...

Page 4: ...2 Date 13 D Units 14 8 1 2 Calibration 14 A Cond TDS 14 A 1 Cell Constant 14 15 A 2 Calibration 15 B Temperature 15 B 1 Status MTC ATC 16 B 2 Temperature Setting 16 b 1 Set Temperature 16 b 2 Temperat...

Page 5: ...uidelines 22 9 2 Troubleshooting Chart 22 Section 10 Maintenance 10 1 General Guidelines 23 10 2 Maintenance Parts 23 10 3 Panel Mounting Parts 23 10 4 Field Mounting Parts 23 Section 11 Technical Dra...

Page 6: ...ature which can be customized as per user selection Press upward and downward button to change the value of digit from 0 9 Press forward button for next digit Press enter button to enter the password...

Page 7: ...2 4 20mA corresponding to temperature 0 1mA accuracy 0 1mA resolution maximum load 500 will be added on request Cell Constants 0 01 20 cm fixed 0 001 19 99 cm freely selectable user defined Relays 1...

Page 8: ...ER CLOSED LOOP SYSTEM DRINKING WATER TEXTILES CHEMICALS WASTE WATER Section 2 Controller Mounting 2 1 Panel Mounting KE CD 1000 Controller can be mounted on control panel plate for different industria...

Page 9: ...xternal supply not provided to work properly WARNING Don t connect AC power cables to the KE CD 1000 controller Important Notes All electrical installations must be supervised by a qualified and respo...

Page 10: ...hows the wiring connection names for RCB conductivity sensor KE CDS 1000 series and RCB conductivity controller KE CD 1000 Each sensor wire connection detail is given below 1 For 2 wire temperature se...

Page 11: ...blue marked connection with negative terminal 4 5 Relay HI LO Connection Two Relay lo Relay hi cables can be connected to the KE CD 1000 controller terminal strip by inserting the wires in RELAY HI RE...

Page 12: ...control through its 6 buttons control panel Figure below shows individual keypad buttons with their respective functions UPWARD TOP DOWNWARD BOTTOM FORWARD RIGHT MENU ENTER ACCEPT ESCAPE CANCEL EXIT...

Page 13: ...CATING CONDUCTIVITY TDS VALUE WITH USER SELECTABLE UNITS TEMPERATURE DISPLAY WITH CELSIUS UNITS UPTO TWO DECIMALS TEMPERATURE COMPENSATION MODE DISPLAY MANUAL MODE MTC AUTOMATIC MODE ATC TIME DISPLAY...

Page 14: ...r defined temperature value from temperature settings see section 8 1 2 As temperature readings effect the accuracy of conductivity measurements controller supports RTD PT 1000 temperature probe KE CD...

Page 15: ...n be selected Settings has four minor sub menus as Display Setting Loop Current Clock Setting and Units A Display Setting User can define customized display interface from this Setting Menu Button Ins...

Page 16: ...ar and can be selected Note Press forward button for 3 secs on main screen to view the loop current for conductivity and temperature B 1 Conductivity User can define initial and final set point values...

Page 17: ...Insert Password Downward Button Settings Enter Button Downward Button Loop Current Enter Button Downward Button Temperature Enter Button Loop current for temperature has two sub menus as 4mA and 20mA...

Page 18: ...forward button to switch between hour and minute Press enter to save the time and escape for main screen C 2 Date User can define customized date from this setting Menu Button Insert Password Downward...

Page 19: ...assword Downward Button Calibration Calibration consists of two sub menus as conductivity TDS and temperature A Cond TDS Conductivity TDS calibration can be done from this sub menu Menu Button Insert...

Page 20: ...ay box Press upward or forward and downward button for Increment and Decrement of 1st digit by 1 up to three digits Press forward button for increment of 3rd digit by 1 upward and downward button will...

Page 21: ...be performed using this feature Menu Button Insert Password Downward Button Calibration Enter Button Downward Button Temperature Enter Button Downward Button Temperature Setting Enter Button Temperatu...

Page 22: ...nu Button Insert Password Downward Button Calibration Enter Button Downward Button Temperature Enter Button Downward Button Temperature Setting Enter Button Downward Button Temperature Coefficient Ent...

Page 23: ...rd and downward button for increment and decrement in main value 1st digit by 1 Press forward button for increment in main Value 2nd digit by 1 If the value goes in 3 digits than forward button will i...

Page 24: ...to reverse from 999 value by pressing downward button Press menu button for increment in decimal value by 0 01 The value can be selected to maximum 999 99 Second is the Factor which is used for settin...

Page 25: ...menu button until the 3rd and 4th digit become 22 Press enter button to save password and controller will return to main screen To input password for main menu access press upward and downward for in...

Page 26: ...es with respect to date time and corresponding temperature Menu Button Insert Password Downward Button Others Enter Button Upward Button Logger Data Logger data can view up to 60 different values for...

Page 27: ...checked check for a possible AC ground loop by temporarily moving the sensor to a sample of solution in a beaker or other container 9 2 Troubleshooting Chart ERROR TYPE ERROR CODE DESCRIPTION TROUBLE...

Page 28: ...buttons connectors and LED can also be replaced if required RCB s corresponding personnel will remain in contact until the issue is resolved 10 3 Panel Mounting Parts Controller is already provided w...

Page 29: ...Technical Drawings 11 1 Controller Dimensions KE CD 1000 controller technical detailed dimensions can be seen below Notes This dimensional drawing is not to scale All dimensions are in mm with inches...

Page 30: ...IS PROVIDED FOR INSULATION OF CONTROLLER WITH ELECTRICAL PANEL AND IP 65 PROTECTION BACK PART INCLUDES POWER SUPPLY PCB MAIN PCB AND CONNECTORS SIDE CLAMP PLATE IS PROVIDED FOR CLAMPING OF CONTROLLER...

Page 31: ..._______________________________________ _________________________________________ _________________________________________ _________________________________________ __________________________________...

Page 32: ...KE CD 1000 PIM R0 2020 RCB Chemicals Services All rights Reserved 24 A 10 SECTOR B1 TOWNSHIP LAHORE PAKISTAN POSTAL CODE 54770 Tel 92 300 064 54 54 Web www rcbchemicals com Email info rcbchemicals com...

Reviews: