12 / 25
Press menu button for increment in decimal values by 0.01.
The value can be selected to
Maximum 999.99. Second is
the
Factor
which is used for
setting
the
value
above
999.99. Factor can be put as 1
or 10 by pressing enter after
selection of factor tab.
Example: If required to set the value of conductivity to
8555 μS/cm
. Firstly check the units if
they are set to
μS/cm
, next set the value to
855.5
from set value and set factor to
10.
This will
make set point value as
855.5 x 10 = 8555
μS/cm
. Press enter to save and escape to return
for set value again. Press escape again for main screen.
Follow the same procedure for
20 mA
option.
B.2 Temperature
User can define initial and final set point values for
Temperature.
Menu Button Insert Password Downward Button
Settings Enter Button Downward Button Loop
Current Enter Button Downward Button
Temperature
Enter Button
Loop current for temperature has two sub menus as 4mA and
20mA.
mA
set point value contains only one input. Upward and
downward button for increment and decrement in main value
by 1, press forward button and menu button for increment and
decrement of decimal digit by 0.1. If the value goes larger than
100 its recommended to come backward from
200 value by pressing downward button. Press enter to save
and escape to return for main screen.
Follow the same procedure for
20 mA
option.
C Clock Setting
Controller contains the feature of displaying date
and time on its main display. Customized date and
time can be set from this setting.
Menu Button Insert Password Downward
Button Settings Enter Button Downward
Button Clock Setting Enter Button
Clock Setting contains Time and Date settings.