16
MagCore® HF48
Please place the Cartridge for test prepared in advance on
the Cartridge Rack for it to be located in the system and insert
thermometer into it after the water of 500µl is put in heat block well
1 (indicated by the arrow)
Select Heater Test in Maintenance menu
Press Start to start test Concurrently, the touch screen shows the
heater temperature tested currently by the system Please compare
to see if the temperature of your thermometer is the same
Please press Stop key if you want to stop test
Maintenance
Please select Item
Back
UV Sterilization
Moving Test
Heater Test
Rejection Tip
Piston Test
O-Ring & Piercing Needles
DC Motor Operate
Update Program Files
━ Heater Test
Start
Stop
Heater Test
Press START to run
Back
Heater A : 65°C / 148°F
Heater B : 65°C / 148°F
This function is mainly for the system to perform rejection after
emergency stop or operation interrupt due to outage etc.
Select Rejection Tip in Maintenance menu
Please select Start to start the process of Rejection Tip After the
process is completed, the system will be rejected to position
automatically
Maintenance
Please select Item
Back
UV Sterilization
Moving Test
Heater Test
Rejection Tip
Piston Test
O-Ring & Piercing Needles
DC Motor Operate
Update Program Files
━ Rejection Tip
Start
Rejection Tip
Press Start to rejection tip
Back
Rejection B Area Tip
Rejection A Area Tip
A
B
D
A
B
E
Next Step
Next Step
heat block well 1
There difference between the
temperature of the thermometer and
that of the heater in the machine.
!