57
56
Secifications (Camera)
Secifications (Thermometer)
Factory defaults
Default Range
Emissivity/Gain
0.95
0.10 to 1.50
in steps of 0.01
Emissivity Table
Free
30 materials
Mode
normal
Hi Alarm
Standard model
900°C (1600°F)
-30 to 900°C
(-25 to 1600°F)
Sub Zero model
500°C (932°F)
-50 to 500°C
(-58 to 932°F)
Lo Alarm
Standard model
-30°C (-25°F)
-30 to 900°C
(-25 to 1600°F)
Sub Zero model
-50°C (-58°F)
-50 to 500°C
(-58 to 932°F)
Offset Adjust
0°C (0°F)
-10 to 10°C
(-18 to 18°F)
Graphic Display
Auto Range
Auto Range
/ Man Range
Cycle Time
0.2 sec
0.1 sec to 300 sec
Data logger
26 points
100 points max.
pre-set with emissivity 0.95
adjustable only via Software
Temp. Range (Standard Model)
- 30 to 900°C (- 25 to 1600°F)
Temp. Range (Sub Zero Model)
- 50 to 500°C (-58 to 932°F)
Display Resolution
0.1°C (0.2°F)
Accuracy (Infrared, Standard Model) ± 0.75% of reading or
at 25°C (77°F) ambient temperature
± 0.75°C (± 1,5°F), whichever is greater
± 2°C (± 4°F) for targets below -5°C (23°F)
Accuracy (Infrared, Sub Zero Model) ± 1% of reading or ± 1°C (2°F),
at 25°C (77°F) ambient temperature
whichever is greater for targets between
- 5 to 500°C (23 to 932°F), ± 1.5°C (3°F)
for targets between -30 to -5°C (-22 to 23°F)
and ± 2°C (4°F) for targets beween
-50 to -30°C (-58 to - 22°F)
Ambient derating
< 0.05K/K or < 0.05%/K,
whichever is greater at
+ 25°C (77°F) ± 25°C (± 45°F)
Optical Resolution
60:1 (19mm spot size at 1.15m.)
(Standard Focus)
(0.75in. spot size at 3.8 feet)
Optical Resolution
50:1( 6mm spot size at 0.3m.)
(Close Focus)
(0.24in. spot size at 0.98 feet)
Accuracy
± 2°C or ± 0.75%,
(Thermocouple K & J)
whichever is greater
Accuracy
(Thermistor)
-30 to
0°C (-22 to 32°F)
± 0.6°C
0 to 70°C (32 to 158°F)
± 0.4°C
70 to 100°C (158 to 212°F)
± 1°C
100 to 120°C (212 to 248°F)
± 1.5°C
Repeatability
± 0.5% of reading or ± 0.5°C (1°F),
(Infrared)
whichever is greater,
± 1°C (± 2°F) for targets
below -5°C (23°F)
Response Time (95%)
250 mSec
Hot Spot Detection (30%)
85 mSec
Spectral Range
8 to 14 µm
Ambient Operating Range
0 to 50°C (32 to 122°F)
Storage Temperature
-20 to 50°C (-4 to 122°F) without batteries
Relative Humidity
10 to 95% at 30°C (86°F), non condensing
Analog output
1 mV/°C(°F)
Digital Output
USB 1.1
Power
2 x 1.5 V Alkaline Type AA
Battery Life (continuous use)
8 hrs (13 hrs with photo mode off)
Dimensions
240 x 170 x 50 mm (7.9 x 6.7 x 2 inches)
Tripod Mount
1/4”-20 UNC
Maximum Picture Number
26
640x480 Pixels (VGA)
Maximum Picture Number
100
320x240 Pixels (1/4 VGA)
Recharge Time for Flash
app. 5 sec
Useful Flash Range
Standard Focus
0.5 to 2m (19 to 79 in.)
Close Focus
0.2 to 1m ( 8 to 40 in)
Camera Lens
6 mm (approx. equal to
42 mm on
a 35 mm camera)
Focal Points
200 mm (8 in.)(Close-up)
Infinity (Far Distance)
Light Sensitivity
6 lux
Shutter Speed
variabel, max. 1/15 sec
Data Interface
USB 1.1
Image File Format