G.Trouble Shooting
System does not display anything:
1. Check the display for any Initialization.
2. Check the fuse. The fuse is located in the Power switch / plug connector.
Please see chapter ‘5.1 Front Panel Overview’. You will find it at position 1.
Touch Panel does not work:
Please connect any USB mouse to the USB port on the front panel
TR-MARK III will display a cursor and everything can be done by mouse. Please see chapter
‘8.5.1.4 Cursor’ if cursor is not automatically displayed.
Measurement can not be started
:
Is the Emergency stop switch pushed in? Turn to release it.
Measurement Accuracy sometimes is not good:
The Raytech TR-MARK III 250V has been designed and tested to be immune against
electromagnetic disturbances. However in rare cases, in presence of strong radio frequency
electromagnetic field, turn ratio measuring accuracy may be degraded depending on field
amplitude and frequency band. In this case, try to keep the instrument away from
disturbances source and/or change cables position. For more information please contact our
local representative or contact Raytech Switzerland.
USB Memory Stick does not work:
There are a few unsupported memory sticks available on the market. Please use another
model and try again.
NOTE
TR-MARK III is designed to be trouble free.
If problems or questions do arise please contact your nearest
representative or our service support group in Switzerland.