Standards and approvals
Safety
IEC 61010-1
EMC
Emission:
Class B
RF emission
CISPR 11
Current harmonic
emission
IEC 61000-3-2
Flicker and voltage
fluctuations
IEC 61000-3-3
Immunity:
industrial
electromagnetic
environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electromagnetic field
IEC 61000-4-3
Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Conducted RF
IEC 61000-4-6
Power frequency
magnetic field
IEC 61000-4-8
Voltage dip
IEC 61000-4-11
Measurement Parameters
Ratio
Resolution: 5 Digits
U
test
Range
Accuracy
250V
0.8 … 5000
±0.06%
Rdg ±1 LSD
5001 … 10’000
±0.1%
Rdg ±1 LSD
10’001 … 32’500
±0.3%
Rdg ±1 LSD
32’501 … 40’000
±0.4%
Rdg ±1 LSD
U
test
Range
Accuracy
100V or 40V
0.8 ... 2000
±0.06%
Rdg ±1 LSD
2001 ... 4000
±0.1%
Rdg ±1 LSD
4001 ... 13000
±0.3%
Rdg ±1 LSD
13000 ... 16000
±0.4%
Rdg ±1 LSD
U
test
Range
Accuracy
10V
0.8 … 500
±0.1%
Rdg ±1 LSD
501 … 1000
±0.12%
Rdg ±1 LSD
1001 … 3250
±0.3%
Rdg ±1 LSD
3251 … 4000
±0.4%
Rdg ±1 LSD
U
test
Range
Accuracy
1V
0.8 … 50
±0.1%
Rdg ±1 LSD
51 … 100
±0.15%
Rdg ±1 LSD
101 … 325
±0.3%
Rdg ±1 LSD
326 … 400
±0.4%
Rdg ±1 LSD
90100-2.20 Instruction Manual TR-MARK III 250V
Page 47 of 68