Ratek SWB10D Water Bath - Operating Manual
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 20 of 24
CAUTION
CAUTION
CAUTION
CAUTION
NEVER
NEVER
NEVER
NEVER operate the water bath if the over-temperature safety cutout test has failed. Operating the water
bath without a functional over-temperature safety cutout may result in serious injury, threat to life, fire or
damage to property.
-
Turn off the mains power supply at the outlet.
-
Allow the unit to cool for at least 10 minutes.
-
Refill the water tank by referring to the section in this User Manual labeled “Filling The Water Tank”.
-
Connect power to the water bath by referring to the section in this User Manual labeled “Filling The
Water Tank”.
-
On the rear panel of the water bath, the Over-temperature reset button (element 11 in Figure 1) should
be pressed in firmly until it remains in place. A small click sound should be heard.
-
If the over-temperature safety cutout test passed, the water bath is now ready to be re-operated as per
the section of this User Manual labeled “Switching On The Water Heating”.