background image

 

16 

 

5. Troubleshooting     

 

Problem 

Possible Causes   

→    Solution 

Sections overlap or 

crumple 

1. Specimen is too warm   

→    Lower specimen temperature   

2. Blade holder too warm   

→      Lower chamber temperature. 

          Tip 

– use freezer spray on clamp plate and blade to confirm.

 

 

3. The anti-roll plate is too low   

→    Raise the anti-roll plate towards 

specimen. 

4. The anti-roll plate and/or clamp plate dirty   

→      Clean with absolute 

alcohol & dry thoroughly. 

Section rolls up under 

the anti-roll plate 

1.Specimen is too cold   

→    Raise specimen temperature. 

Specimen curls after 

lifting the anti-roll 

plate 

1. Anti-Roll plate & clamp plate too warm   

→    Lower chamber temperature 

Tip 

– use freezer spray on clamp plate and blade to confirm.

 

2. Static electricity in chamber         

3. Blade blunt   

→      Change blade. 

Sections tear or crack 

1. Specimen too cold   

→      Raise temperature.   

2. Blade damaged or dirty   

→      Change blade. 

3. Specimen frozen too rapidly or specimen overly large. 

Specimen and 

section 

chatter 

1. Blade holder not correctly clamped   

→    Check & tighten blade holder. 

2.  Blade  incorrectly  clamped   

→  Blade  clamping  force  can  be  increased  by 

tightening clamp screw at the rear of the top stage. 

3.  Specimen  incorrectly  clamped   

→    Check  specimen  is  securely  mounted 

and  clamped  in  specimen  head  jaws.  Check  there  is  no  debris  or  ice  on  the 

back of the chuck or on specimen head. 

Tissue sections stick 

to the anti-roll plate 

1.  The  anti-roll  plate  is  dirty   

→    Use  a  short  brush  to  clean  the  anti-rolling 

plate. 

2. The plate is too warm   

→    Prolong the refrigeration time. 

Vertical crack of 

tissue section 

Edge of the blade has a defect or has a foreign object   

→    use a new blade or 

remove the object gently with a brush 

Sections thick-thin 

1. Check blade holder and specimen correctly clamped. 

2. Ensure specimen is securely attached to the cassette/chuck. 

3. Temperature of specimen incorrect   

→    Raise or lower temperature. 

4. Blade not sharp   

→      Change blade. 

Vibration when slicing 

1. Check blade holder and specimen correctly clamped. 

2. The blade angle is too high or too low. 

3. Ensure specimen is securely attached to the specimen disc/chuck. 

4. Sectioning speed is too fast . 

 

 

Summary of Contents for Basics CRY85

Page 1: ...ttention Before using the instrument please read this Instruction carefully and keep the Instruction for further reference Please refer to the nameplate on the back of the instrument for the serial nu...

Page 2: ...t 6 Middle Peltier freezing 7 Middle Standby 7 Middle Sectioning trimming setting 7 Right Clock setting 8 Right Temperature setting 8 Right Fast freezing 8 Right Start of work timer 9 Right End of wor...

Page 3: ...oning procedures 14 Tips for Successful Sectioning 14 Recommended Temperature Ranges 15 5 Troubleshooting 16 6 Cleaning and Maintenance 17 Cleaning the display screens 17 Cleaning procedures 17 Empty...

Page 4: ...men head cooling UV option enables surface disinfection of harmful bacteria viruses and spores by using 35 min after each use The Peltier semiconductor cooling function on the quick freeze shelf can b...

Page 5: ...50 m Specimen retraction Adjustable range 0 80 m Increment 5 m Horizontal stroke 20 mm Vertical stroke 60 mm Voltage AC 110V 10 AC 220V 10 Frequency 50 60Hz Power 600W Largest start up current 5 secon...

Page 6: ...s as well as the back of the instrument to facilitate ventilation and heat dispersion The instrument can be used after it is kept standing still for at least 2 hours Please ensure the grounding wire o...

Page 7: ...e To protect the instrument from being damaged by an abrupt startup the instrument cannot be re started within 3 minutes after it is turned off The central operation panel includes three zones left mi...

Page 8: ...utes which will count down to 14 13 2 1 until the Peltier stops cooling and the indicator light is off User may also press this button to stop the Peltier when it is running The Peltier stops running...

Page 9: ...actual temp Peltier on off Clock setting The clock can be set by pressing D to adjust weekday and pressing H and M to adjust hour and minute respectively Temperature setting Press the temperature but...

Page 10: ...rking day then press the H and M buttons to adjust hour and minute respectively The system will automatically set 24 hours of the standby mode for non working days Programming end of work timer Press...

Page 11: ...le by moving forward Press button specimen is moving towards user slowly Fast Advance and Retract Press double arrow buttons to move the specimen head forward or backward The indicator light on the bu...

Page 12: ...lly move the holder After the holder is placed in the desired position lock the lever clockwise Note Please lock the blade angle in a proper position before adjusting forward backward movements Blade...

Page 13: ...d in place by two screws Ensure that the gap clearance side of the anti roll plate is facing inwards Position the anti roll plate in the holder gently pushing the glass to the bottom of the holder The...

Page 14: ...eel Specimen Setup The specimen must be correctly secured and oriented in the specimen head for subsequent sectioning To secure and orient the specimen Loosen the specimen chuck fixation screw and ins...

Page 15: ...late and the clamping plate i e with ethanol Tighten all clamping screws and clamping levers on the blade holder and specimen head Select the appropriate knife carrier and specimen temperature accordi...

Page 16: ...t 35 C to 40 C 30 C or below Cervix 25 C to 30 C 20 C Connective Tissue 25 C 16 C Gut 25 C 20 C Heart 25 C 20 C Kidney 25 C 15 C Lip 25 C 13 C Liver 25 C 13 C Lung 25 C 15 C Lymph Node 25 C 13 C Muscl...

Page 17: ...holder not correctly clamped Check tighten blade holder 2 Blade incorrectly clamped Blade clamping force can be increased by tightening clamp screw at the rear of the top stage 3 Specimen incorrectly...

Page 18: ...lade holder Carefully clean and dry the dismounted blade holder As there is condensate humidity inside the microtome structure dry the microtome components inside the cryochamber very carefully For th...

Page 19: ...ion of the instrument If the original packaging is no longer available please contact customer service via support rankinbiomed com II The conditions for storage and transportation must be observed du...

Page 20: ...this instrument is designed to accepted standards of safety Its use does not entail any hazard if operated in accordance with the instructions given in the documentation However the following safety p...

Page 21: ...Item Description Qty Special Note 1 Base Unit 1 2 Disposable Blade Holder 1 2 in 1 low and high profile 3 Handwheel Handle 1 4 Specimen Disc 3 5 Power Cord 1 6 Brush 2 7 Heated Glass Door 1 Removable...

Page 22: ...ers dealers or resellers Limitation of Warranty The warranty stated above is the only warranty applicable to this product All other warranties expressor implied including all implied warranties of mer...

Reviews: