4.2 Side operation panel 

 

 

The panel on the left of the cryochamber controls the specimen head feeding, cryochamber 

illumination, and UV sterilizing.   

 

 

 

             

UV sterilization 

Press this UV sterilization button to turn on the UV lamp and sterilize the cryochamber, as 

indicated by the indicator light on the upper left corner of the button. The sterilization will be 

turned  off  automatically  after  35  minutes  or  manually  by  pressing  this  button  again  any 

time. 

 

 

 

 

 

Cryochamber illumination   

 

Press this button to turn on/off the light inside the cryochamber.   

 

Fine Advance   

Specimen position is adjustable by moving forward. Press 

” button, specimen is 

moving towards user slowly. 

 

Fast Advance and Retract 

   

Press  double  arrow  buttons  to  move  the  specimen  head  forward  or  backward.  The 

indicator light on the button will turn on and alarm will beep when reaching the front or 

rear limit. When the fast retract button is pressed, the specimen head will return all the 

way to the rear limit.   

 

                                           

Note

 

1.  Please  keep  the  quartz  tube  clean,  which  is  critical  for  successful 

sterilization. The sterilization should be conducted under the standby 

mode rather than the super-low temperature condition. 

2.  Please  close  the  chamber  glass  window  during  UV  sterilization  to 

ensure the effectiveness of sterilization. 

 

Note

When adjusting the forward or backward movement of the 

specimen, the hand wheel must be locked. 

Summary of Contents for Basics CRY85

Page 1: ...ttention Before using the instrument please read this Instruction carefully and keep the Instruction for further reference Please refer to the nameplate on the back of the instrument for the serial nu...

Page 2: ...t 6 Middle Peltier freezing 7 Middle Standby 7 Middle Sectioning trimming setting 7 Right Clock setting 8 Right Temperature setting 8 Right Fast freezing 8 Right Start of work timer 9 Right End of wor...

Page 3: ...oning procedures 14 Tips for Successful Sectioning 14 Recommended Temperature Ranges 15 5 Troubleshooting 16 6 Cleaning and Maintenance 17 Cleaning the display screens 17 Cleaning procedures 17 Empty...

Page 4: ...men head cooling UV option enables surface disinfection of harmful bacteria viruses and spores by using 35 min after each use The Peltier semiconductor cooling function on the quick freeze shelf can b...

Page 5: ...50 m Specimen retraction Adjustable range 0 80 m Increment 5 m Horizontal stroke 20 mm Vertical stroke 60 mm Voltage AC 110V 10 AC 220V 10 Frequency 50 60Hz Power 600W Largest start up current 5 secon...

Page 6: ...s as well as the back of the instrument to facilitate ventilation and heat dispersion The instrument can be used after it is kept standing still for at least 2 hours Please ensure the grounding wire o...

Page 7: ...e To protect the instrument from being damaged by an abrupt startup the instrument cannot be re started within 3 minutes after it is turned off The central operation panel includes three zones left mi...

Page 8: ...utes which will count down to 14 13 2 1 until the Peltier stops cooling and the indicator light is off User may also press this button to stop the Peltier when it is running The Peltier stops running...

Page 9: ...actual temp Peltier on off Clock setting The clock can be set by pressing D to adjust weekday and pressing H and M to adjust hour and minute respectively Temperature setting Press the temperature but...

Page 10: ...rking day then press the H and M buttons to adjust hour and minute respectively The system will automatically set 24 hours of the standby mode for non working days Programming end of work timer Press...

Page 11: ...le by moving forward Press button specimen is moving towards user slowly Fast Advance and Retract Press double arrow buttons to move the specimen head forward or backward The indicator light on the bu...

Page 12: ...lly move the holder After the holder is placed in the desired position lock the lever clockwise Note Please lock the blade angle in a proper position before adjusting forward backward movements Blade...

Page 13: ...d in place by two screws Ensure that the gap clearance side of the anti roll plate is facing inwards Position the anti roll plate in the holder gently pushing the glass to the bottom of the holder The...

Page 14: ...eel Specimen Setup The specimen must be correctly secured and oriented in the specimen head for subsequent sectioning To secure and orient the specimen Loosen the specimen chuck fixation screw and ins...

Page 15: ...late and the clamping plate i e with ethanol Tighten all clamping screws and clamping levers on the blade holder and specimen head Select the appropriate knife carrier and specimen temperature accordi...

Page 16: ...t 35 C to 40 C 30 C or below Cervix 25 C to 30 C 20 C Connective Tissue 25 C 16 C Gut 25 C 20 C Heart 25 C 20 C Kidney 25 C 15 C Lip 25 C 13 C Liver 25 C 13 C Lung 25 C 15 C Lymph Node 25 C 13 C Muscl...

Page 17: ...holder not correctly clamped Check tighten blade holder 2 Blade incorrectly clamped Blade clamping force can be increased by tightening clamp screw at the rear of the top stage 3 Specimen incorrectly...

Page 18: ...lade holder Carefully clean and dry the dismounted blade holder As there is condensate humidity inside the microtome structure dry the microtome components inside the cryochamber very carefully For th...

Page 19: ...ion of the instrument If the original packaging is no longer available please contact customer service via support rankinbiomed com II The conditions for storage and transportation must be observed du...

Page 20: ...this instrument is designed to accepted standards of safety Its use does not entail any hazard if operated in accordance with the instructions given in the documentation However the following safety p...

Page 21: ...Item Description Qty Special Note 1 Base Unit 1 2 Disposable Blade Holder 1 2 in 1 low and high profile 3 Handwheel Handle 1 4 Specimen Disc 3 5 Power Cord 1 6 Brush 2 7 Heated Glass Door 1 Removable...

Page 22: ...ers dealers or resellers Limitation of Warranty The warranty stated above is the only warranty applicable to this product All other warranties expressor implied including all implied warranties of mer...

Reviews: