![R&S VTE Getting Started Download Page 20](http://html1.mh-extra.com/html/rands/vte/vte_getting-started_775062020.webp)
Preparing for Use
R&S
®
VTE
20
Getting Started 2116.1270.02 ─ 07
b) Mount the adapter kit. Follow the assembly instructions provided with the
adapter kit.
2. Lift the product to shelf height.
3. Grab the handles and push the product onto the shelf until the rack brackets
fit closely to the rack.
4. Tighten all screws at the rack brackets to secure the product at the rack.
To unmount the product from a rack
1. Loosen the screws at the rack brackets.
2. Remove the product from the rack.
3. If placing the product on a bench top again, unmount the adapter kit from the
product. Follow the instructions provided with the adapter kit.
4.4
Considerations for Test Setup
Cable selection and electromagnetic interference (EMI)
Electromagnetic interference (EMI) can affect the measurement results.
To suppress electromagnetic radiation during operation:
●
Use high-quality shielded cables, especially for connecting the following:
– RF/marker/trigger/reference inputs and outputs: Use cables with at least
80
dB to 1
GHz shielding, usually achieved by double-shielded cables.
– USB interfaces: Use double-shielded cables no longer than 1
m. Use only
USB devices that remain within the permissible EMI limits.
– USB device port: Use double-shielded cables no longer than 5
m.
– Display port interface: Only use peripheral equipment that does not cause
limit violations.
– DVI interface: Use a cable shielded with ferrite cores.
– LAN interface: Use category
7 cables.
●
Always terminate open cable ends.
●
Ensure that connected external devices comply with EMC regulations.
Considerations for Test Setup
Summary of Contents for VTE
Page 1: ...R S VTE Video Tester Getting Started Getting Started Version 07 2116127002 E Ô2 ...
Page 6: ...Contents R S VTE 6 Getting Started 2116 1270 02 07 ...
Page 16: ...Key Features R S VTE 16 Getting Started 2116 1270 02 07 ...
Page 26: ...Preparing for Use R S VTE 26 Getting Started 2116 1270 02 07 Turn On Tests ...
Page 42: ...Operating the R S VTE in a LAN R S VTE 42 Getting Started 2116 1270 02 07 Firewall Settings ...
Page 50: ...Contacting Customer Support R S VTE 50 Getting Started 2116 1270 02 07 ...