Remote Commands Reference
R&S
®
RTC1000
236
User Manual 1335.7352.02 ─ 02
Table 13-3: Header data
Position
Meaning
Example
1
XStart in s
-9.477E-008 = - 94,77 ns
2
XStop in s
9.477E-008 = 94,77 ns
3
Record length of the waveform in samples
200000
4
Number of values per sample interval, usually 1.
1
Suffix:
<m>
.
1..4
Selects the reference waveform, the internal reference storage.
Parameters:
<Header>
Comma-separated value list
Example:
-9.477E-008,9.477E-008,200000,1
Usage:
Query only
REFCurve<m>:DATA:XINCrement?
Returns the time difference between two adjacent samples of the indicated waveform.
The commands are relevant for data conversion if binary data format is defined (FORM
UINT, 8 | 16).
Suffix:
<m>
.
1..4
Selects the reference waveform, the internal reference storage.
Return values:
<Xincrement>
Time in s
Usage:
Query only
REFCurve<m>:DATA:XORigin?
Returns the time of the first sample of the indicated waveform. The commands are rel-
evant for data conversion if binary data format is defined (FORM UINT, 8 | 16).
Suffix:
<m>
.
1..4
Selects the reference waveform, the internal reference storage.
Return values:
<Xorigin>
Time in s
Usage:
Query only
REFCurve<m>:DATA:YINCrement?
Returns the voltage value per bit of the indicated waveform. The commands are rele-
vant for data conversion, if binary data format is defined (FORM UINT, 8 | 16).