Serial Bus Analysis
R&S
®
RTC1000
148
User Manual 1335.7352.02 ─ 02
11.7.1 LIN Bus Configuration
Before you configure the bus, set the correct logic level (threshold).
●
●
"To set the threshold for logic states"
default value is 500
mV.
To select the LIN bus
1. Press the BUS key in the Vertical section.
2. In the short menu, select the bus: "B1" or "B2".
3. In the "BUS" menu, press the "BUS TYPE" softkey.
4. Use the UNIVERSAL knob to select "LIN":
To configure the CAN bus
1. In the "BUS" menu, press "CONFIGURATION".
2. Configure the bus. The settings are described below.
Figure 11-19: LIN bus configuration
LIN Bus (Option R&S
RTC-K3)