Remote Control
R&S
®
NGE100B
84
User Manual 5601.1343.02 ─ 04
FUSE:LINK
<arg0>
This command defines or queries the channel fuses (fuse linking) for the previous
selected channel.
Parameters:
<arg0>
{1 | 2 | 3}
1
channel Ch 1
2
channel Ch 2
3
channel Ch 3
Example:
INST OUT1
FUSE:LINK 2
Fuse Ch 1 is linked with Ch 2
FUSE:LINK? 2
-> 1
If the fuse of Ch 1 is linked with Ch 2, a "1" is returned.
FUSE:LINK? 3
-> 0
If the fuse of Ch 1 is not linked with Ch 3, a "0" is returned.
FUSE:TRIPped?
This command queries the fuse tripped of the previous selected channel.
Return values:
<arg0>
ON
Fuse is tripped.
OFF
Fuse is not tripped.
Example:
INST OUT1
FUSE:TRIP?
-> 0
Fuse of Ch 1 has not tripped.
Usage:
Query only
FUSE:UNLink
<arg0>
This command unlinks the channel fuses (fuse linking) for the previous selected chan-
nel.
Parameters:
<arg0>
{1 | 2 | 3}
1
channel Ch 1
2
channel Ch 2
Remote Control - Commands