![background image](http://html1.mh-extra.com/html/rands/nge100b-series/nge100b-series_user-manual_3083288093.webp)
Remote Control
R&S
®
NGE100B
90
User Manual 5601.1343.02 ─ 04
Example:
INST OUT1
APPLY "6,2"
Set channel 1 to output 6 V and 2 A.
APPLY?
-> 6, 2
Return voltage and current value of the selected channel.
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
<Current>
This command defines or queries the current value of the selected channel.
Parameters:
<Current>
<Current>
Set the selected channel to a current range of 3 mA to 3.025 A
Default unit: A
Example:
CURR 2
Set the selected channel current to 2 A.
CURR?
-> 2
Return current value of the selected channel .
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
<Voltage>
This command defines or queries the voltage value of the selected channel.
Parameters:
<Voltage>
{<Voltage> | MINimum | MAXimum | DEFault}
<Voltage>
0.0 V to 32.2 V (adjustable in 10 mV steps)
MINimum
0.000E+00
MAXimum
32.200E+01
Default unit: V
Example:
VOLT 10
Set the voltage value to 10 V.
VOLT?
-> 10
Return the voltage value of the selected channel.
6.5.12
EasyRamp Commands
[SOURce:]VOLTage:RAMP:DURation
<Time>
This command sets/queries the EasyRamp duration of the previous selected channel
in seconds. 0.01 s to 10 s in steps of 0.01 s.
Remote Control - Commands