Remote Commands
R&S
®
FSMR3000
429
User Manual 1179.0116.02 ─ 02
Parameters:
<Frequency>
The allowed range and f
max
is specified in the data sheet.
*RST:
fmax/2
Default unit: Hz
Example:
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Sets the center frequency to 110 MHz.
Manual operation:
See
[SENSe:]FREQuency:CENTer:STEP
<StepSize>
This command defines the center frequency step size.
Parameters:
<StepSize>
f
max
is specified in the data sheet.
Range:
1 to fMAX
*RST:
0.1 x span
Default unit: Hz
Example:
//Set the center frequency to 110 MHz.
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Manual operation:
See
[SENSe:]FREQuency:CENTer:STEP:LINK
<CouplingType>
This command couples and decouples the center frequency step size to the span or
the resolution bandwidth.
Parameters:
<CouplingType>
SPAN | RBW | OFF
SPAN
Couples the step size to the span. Available for measurements
in the frequency domain.
OFF
Decouples the step size.
*RST:
SPAN
Example:
//Couple step size to span
FREQ:CENT:STEP:LINK SPAN
Configuring the Measurement