Remote Commands
R&S
®
FSMR3000
401
User Manual 1179.0116.02 ─ 02
12.5.1.1
RF input
INPut<ip>:ATTenuation:PROTection:RESet
.......................................................................401
INPut<ip>:ATTenuation:PROTection[:STATe]
.....................................................................401
INPut<ip>:FILTer:HPASs[:STATe]
..................................................................................... 402
..........................................................................................403
INPut<ip>:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
for the R&S
FSMR3000 after an overload condition occurred and the protection mech-
anism intervened. The error status bit (bit 3 in the
STAT:QUES:POW
status register)
and the
INPUT OVLD
message in the status bar are cleared.
The command works only if the overload condition has been eliminated first.
Suffix:
<ip>
.
1 | 2
irrelevant
Example:
INP:ATT:PROT:RES
INPut<ip>:ATTenuation:PROTection[:STATe]
<State>
This command turns the availability of attenuation levels of 10
dB or less on and off.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
ON | 1
Attenuation levels of 10
dB or less are not allowed to protect the
RF input connector of the R&S
FSMR3000.
OFF | 0
Attenuation levels of 10
dB or less are not blocked. You must
provide appopriate protection for the RF input connector of the
R&S
FSMR3000 yourself.
*RST:
1
Example:
INP:ATT:PROT ON
Turns on the input protection.
INPut<ip>:COUPling
<CouplingType>
This command selects the coupling type of the RF input.
Configuring the Measurement