OPERATION
2 - 14
2.6 Preset Alarm Limits and Calibration
The UltraRAE Compound Specific PID monitor is factory
calibrated with standard calibration gas, and is programmed
with default alarm limits as listed below. Refer to Section 4 for
programming procedures if new calibration or alarm limits are
required.
Factory Calibration and Preset Alarm Limits
Cal Gas
Cal Span Units
Low
High
Benzene
5.0
ppm
5.0
10.0
Butadiene
5.0 ppm
5.0
10.0
Halocarbon
10.0
ppm 10.0 25.0
VOC
100.0
ppm 50.0 100.0