GENERAL INFORMATION
1 - 4
Intrinsic Safety:
UL & cUL
Class 1, Division I,
Groups A, B, C, D (US & Canada)
0575 II 1G
DEMKO 03 ATEX 0204759X
Eex ia IIC T4 (Europe)
EM Interference:
no effect when exposed to 0.43 mW/cm2
RFI (5 watt transmitter at 12 inches)
Alarm Setting:
preset alarm limits for two action levels
Audible Alarm:
90 dB buzzer @ 10 cm
Visual Alarm
:
flashing red LED
External Alarm:
optional plug in pen size vibration alarm
or remote alarm
Real-time Clock:
automatic date and time stamps data
logged information
Data logging:
4000 data points, includes serial number,
user ID and site ID, date and time
Communication:
down load data to PC and upload
instrument setup from PC through RS-232
link to serial port on PC
Calibration:
two-point field calibration of zero and
standard reference gas for compound
specific vapor
Sampling Pump:
internal integrated flow rate, 400 cc per
minute minimum (VOC tube), low flow
alarm
Temperature:
-20o to 45oC (-4o to 113oF)
Humidity:
0 % to 95 % relative humidity
(Non-condensing)
Housing:
ABS + PC, conductive coating, splash and
dust proof, will withstand 10’ drop test
Attachment:
wrist strap