ToxiRAE Pro User’s Guide
43
17.
When you see this confirmation. Click “Yes.”
Uploading will take a few seconds, and this progress bar is shown:
18.
When the upload is done, exit ProRAE Studio II.
19.
Press [Y/+] on the ToxiRAE Pro to exit Communication Mode.
14. Bump Testing & Calibrating The ToxiRAE Pro
Note:
Calibration intervals and bump test procedures may vary due to national
legislation.
14.1. Bump Testing & Calibrating With An AutoRAE 2™
Bump testing and calibrating a ToxiRAE Pro can be done with an AutoRAE 2 Automatic
Test And Calibration Station. Follow all instructions in the AutoRAE 2 User’s Guide for
successful bump testing or calibration.
14.1.1. Bump (Functional) Testing
With the ToxiRAE Pro in Normal Mode:
1.
Connect the calibration gas cylinder, flow regulator, and calibration adapter to the
ToxiRAE Pro and start the gas flow.
2.
Make sure the unit goes into alarm and that the buzzer produces at least two beeps per
second, the LED lights flash on and off, and the vibration alarm functions. The
display backlight should illuminate and an alarm message should be shown in the
display.
3.
Turn off the gas flow.
4.
Remove the calibration adapter.